![Optimanufacturing T60-1001-C8-1A Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/optimanufacturing/t60-1001-c8-1a/t60-1001-c8-1a_operation-and-service-manual_1642723010.webp)
SOP-TN-002 Rev. Q
T60 Operations and Service Manual
9
SECTION V
PERIODIC INSPECTION, MAINTENANCE, AND LUBRICATION
5-1.
GENERAL -
No periodic inspection, maintenance, or lubrication is required for cable Tensiometers.
As a matter of general practice, the instrument should be kept clean at all times. Particular attention
should be paid to the two sectors and the riser block to see that they are always free from foreign
matter. The instrument as shipped by the manufacturer has been lubricated. In the event that repairs
are made as outlined in Section VI, lubrication should be accomplished as listed in table 3. Method of
application is a light wipe.
*Explanation: D. C. #33: Dow Corning Grease #33 fluid type or equivalent to NATO G-395 fluid type per MIL-
PRF-81322.
Table 3- Lubrication Chart
FIG. 6
INDEX NO.
LUBRICATION*
LOCATION
20
D.C. #33
Both sides
23
D.C. #33
On OD of pins and flat face
26
D.C. #33
ID and OD
29
D.C. #33
OD
32
D.C. #33
OD
38
D.C. #33
OD
40
D.C. #33
OD
SECTION VI
TROUBLE SHOOTING
6-1.
GENERAL
-
The instrument will be free from trouble such as would be caused by nominal use. The
instrument, however, can be damaged by mistreatment and improper usage. If a detail part of the
instrument becomes damaged or broken, disassemble the instrument in the order of the index
numbers indicated on the exploded view, Figure 6. All detail parts of the instrument are so integrated
that in order to perform a disassembly operation it is necessary that disassembly be started with index
number 1 and continued in numerical sequence. The only exception to this is the latch mechanism (37
through 40, Figure 6). Reassembly is the opposite of disassembly. Whenever the instrument has been
disassembled, it must be recalibrated. (See Section VII.)
6-2.
SERVICE TROUBLES AND REMEDIES -
(See table 4) a listing of common troubles together with
their probable causes and remedies is presented as a trouble-shooting chart, table 4.
Table 4- Trouble Shooting Chart
TROUBLE
PROBABLE CAUSE
(See Figure 6)
REMEDY
Tensiometer reads high
Faulty dial indicator (35)
Replace dial indicator and calibrate
Tensiometer reads low
Weak trigger spring (29)
Replace trigger spring and calibrate
Dirt and foreign matter between frame
(41) and sectors (24)
Clean and reassemble
Worn out sectors (24) or riser Block (5)
Replace sectors and riser block; then
calibrate
Faulty dial indicator (35)
Replace dial indicator and calibrate
Pointer lock will not function
Switch spring (30) broken
Replace switch spring and calibrate
Handle latch will not lock
handle in close position
Handle latch spring (39) weak or
broken
Replace latch spring. Do not calibrate