9. No warning setup: no warning  when there is no loading on the balance  or the 

weight is lower than the lowest weight limit. 
1. Press MENU key.
2. Press TARE key repeatedly until it reads “InSPCT’. 
3. Press COUNT/ENTER key, the display reads ‘SET HI’.
4. Press TARE key repeatedly until it reads ‘nOnrES’.
5. Press COUNT/ENTER key, the display reads “50” (the initialization value which 

indicates  no  warning when the  real weight  is  lower  than 50%  of  the lowest 
weight).

6. Press the COUNT/ENTER key to increase the value, while press TARE key to 

decrease  the  value.  Choose  the  desired  value,  press  MENU  key  then  the 
desired value blinks. Press COUNT/ENTER key to confirm, the display  reads 
“SET HI”.

10.Enabling the inspect setup: to start inspect setup:

1. Press TARE key repeatedly until the display reads “EnAbLE”.
2. Press  COUNT/ENTER  key  to  confirm;  to  disable inspect  setup,  press TARE 

key repeatedly until the display reads “dISAbL”.

3. Press COUNT/ENTER key to confirm, the balance returns to weighing mode.

11.Checking the highest and the lowest weight limits:

1. Press MENU key, the display reads “PrINT”, press TARE key repeatedly until it 

reads “InSPCT”.

2. Press COUNT/ENTER key, the display reads “SET HI”.
3. Press COUNT/ENTER the displays reads the highest weight limit.
4. Press TARE key, the display returns to reads “SET HI”.
5. Press TARE  key, the  display  reads “SET  LO”, press  COUNT/ENTER  key, the 

display reads the lowest weight limit.

Notice:

 To escape the menu setup anytime during the operation, press TARE key until the 

display reads “ESC”, press COUNT/ENTER to confirm.

 To clear the highest and the lowest  weight  limits, press TARE key repeatedly until 

the  display  reads “CLEAr”,  press  COUNT/ENTER  key,  then  the  limits  value  both 
return to zero.

 After  any  modification  of  the highest  and  the  lowest  weight  limits, the  balance 

needs to enter inspect mode again. Follow steps 7 to10 to enable inspect mode.

 The highest and the lowest  weight limits would be saved in the balance. There is 

no need to setup the limits for the next start-up of the balance.

 

OPD-A Series

 

User Manual / Using Scale

 

9/15

Содержание OPD-A Series

Страница 1: ...OPD A Series User Manual ...

Страница 2: ...rial onto the platform as that could damage the balance Do not spill liquids on the balance as that could damage the balance Material that has a static charge could influence the performance of the balance Electromagnetic interference can affect the reading of the balance Unplug the power before cleaning and use a piece of soft cloth do not use any aggressive cleaning agent Make sure no liquid ent...

Страница 3: ...unting Function 7 Percent Deviation Function 8 Inspection Function 8 Print Setup 10 Baud Rate Setup 10 Enabling Units of Measure 10 Restoring Factory Default 11 Backlight Setup 11 Additional Functions 12 RS232 Communication 13 More Info 15 Troubleshooting 15 Technical Support 15 OPD A Series User Manual Table of Contents 3 15 ...

Страница 4: ... of models this manual covers Model Max Weight grams Accuracy grams OPD A103 120 0 001 OPD A203 210 0 001 OPD A303 310 0 001 OPD A403 410 0 001 OPD A503 510 0 001 OPD A1002 1200 0 01 OPD A2002 2100 0 01 OPD A3002 3100 0 01 OPD A4002 4100 0 01 OPD A5002 5100 0 01 OPD A Series User Manual Getting Started 4 15 ...

Страница 5: ...t airflow contact with water and strong electromagnetic wave 4 Level the scale by using the bubble level indicator if available and adjusting the feet of the scale 5 Locate scale in the following condition 1 Temperature 5oC 40oC 2 When relocating the scale warm up the scale for 30 minutes 6 Plug in the scale to a surge protected power outlet 7 Press the ON OFF key in the front panel to turn the sc...

Страница 6: ...le display will show its full range Press the TARE key and the calibration point will change on the display 4 Choose a calibration point place the correct weight on the pan 5 Press the CAL key and the scale should read CAL 6 When the calibration is completed the scale should read the value of the weight on the pan 7 Test the calibration by testing different known weights 8 Re calibrate if measurem...

Страница 7: ... pennyweight 1gram 0 64301493100dwt g gram Counting Function To perform counting function follow the procedure below 1 Place the container on the pan if applicable 2 Press theTARE key 3 Press the COUNT ENTER key to choose the number of reference samples 4 Place the selected number of reference samples on the pan or container if applicable 5 Press the UNIT key the scale should read the number of sa...

Страница 8: ...T 3 Press the COUNT ENTER key to confirm then the scale enters limits setup The display should read SET HI to set the highest weight limit 4 Press the COUNT ENTER key to confirm and the display shows 50 initialize value 5 Press the COUNT ENTER key and the display reads SET dP to set the decimal position PressTARE key repeatedly to choose the decimal position 6 Press the COUNT ENTER key to confirm ...

Страница 9: ... key to confirm the balance returns to weighing mode 11 Checking the highest and the lowest weight limits 1 Press MENU key the display reads PrINT press TARE key repeatedly until it reads InSPCT 2 Press COUNT ENTER key the display reads SET HI 3 Press COUNT ENTER the displays reads the highest weight limit 4 PressTARE key the display returns to reads SET HI 5 Press TARE key the display reads SET L...

Страница 10: ...ess Tare key it reads the other check types Odd indicates odd check and EvEn indicates even check 7 Choose the proper check type and press Count Enter to confirm the balance now returns to weighing mode Enabling Units of Measure The units function can be programmed to turn certain weighing units on or off To enable or disable certain units of measure perform the following procedure 1 Press Menu ke...

Страница 11: ...efault setup is finished Backlight Setup Follow the procedures below 1 Press Menu key the display reads PrInT 2 Press Tare key repeatedly until it reads bLgHT press Count Enter to confirm the display reads 1 nIn which indicates the backlight will be off in 1 minute 3 Press Tare key repeatedly to choose the desired time 1 2 3 5 10 15 30 60 minutes are available to choose from 4 Press Count Enter ke...

Страница 12: ...r example If necessary set a safety guard to avoid air current Density Measurement Purpose This function is used to test the density of solid object with bottom weighing and buoyancy theory Application of Formula Sample Density Liquid Density Mass in Air Mass in Air Mass in Water As the liquid density varies with different temperatures please measure the liquid temperature with the temperature met...

Страница 13: ...2 interface The following commands are available U Unit units conversion T Tare deduct the tare weight C Cal calibration with external standard weight P Print print function percent function instant print M Count counting function When a balance is connected to a computer it is suggested that immediate print be used In response to this command the balance will transmit whatever number or message a...

Страница 14: ...rresponds to the OK indicator on the display S means the reading is stable space means the reading is not stable L retur n char acter M line feeds character it indicates the line feeds RS232 port Parameters Data format 10 bit data frame bit1 bit2 bit 10 bit1 start bit bit 2 9 data bits bit 10 stop bit Baud rate 9600 bps All data in ASCII code Physical pinout Pin 2 TXD Transmit Pin 3 RXD Receive PI...

Страница 15: ... calibration procedure has been performed correctly It won t hurt to run through the calibration process again when in doubt Q The scale does not turn on A Make sure the power cord is plugged in and that there is power One easy way to test this is by connecting another appliance to the same outlet and see if it s operational Technical Support Please email support optimascale com if you have any fu...

Отзывы: