Function
Parameter Settings/Options
Negative Display
C25
0 = -9d
10 = -10% max. capacity
20 = -20% max. capacity
50 = -50% max. capacity
100 = -100% max. capacity
Standstill Time
C26
0 = quick
1 = medium
2 = slow
Standstill Range
C27
1 = 1d d = division
2 = 2d
5 = 5d
10 = 10d
Digital Filter
(for filtering moving
weight such as
animals)
C28
0 = turn off dynamic filter
1 = 1 digital filter strength
2 = 2 digital filter strength
3 = 3 digital filter strength
4 = 4 digital filter strength
5 = 5 digital filter strength
6 = 6 digital filter strength
Note: The higher the number, the higher the filter strength
Noise Filter
C29
0 = turn off noise filter
1 = 1 digital filter strength
2 = 2 digital filter strength
3 = 3 digital filter strength
Gravity of Calibration
Location
C36
9.7000 - 9.9999
Gravity of Destination
C37
9.7000 - 9.9999
Version No.
C38
Print Mode
C41
0 = auto mode
1 = gross mode
2 = tare mode
Print Carriage Return
C42
0 - 9 (How much space between print outs)
Space Print
C43
0 - 9 (Where the data prints on the paper: 0 = left ; 9 = right)
Date Print
C44
0 = do not print the date
1 = print the date
Time Print
C45
0 = do not print the time
1 = print the time