OP-907 SERIES INDICATOR
USER’S MANUAL
(OP-907A, OP-907B Series)
Страница 1: ...optimascale com OP 907 SERIES INDICATOR USER S MANUAL OP 907A OP 907B Series...
Страница 2: ...et Up 1 Features 2 Specifications 3 Power Supply 4 Displays 5 Display and Key Descriptions 6 Operating Instructions 7 9 Calibration 10 11 Indicator Parameter Settings 12 14 Definitions 15 Connectors 1...
Страница 3: ...as it could damage the load cell inside Material that has a static electric charge could influence the weighing Discharge the static electricity of the samples if possible Another solution to the pro...
Страница 4: ...l Parameters Accuracy class 3000 e Stimulating voltage 3 3 VDC A D converting speed 10 SPS Load signal range 0 12 8mV Load capacity can connect 4 pcs 350 load cell at most Interval 1 2 5 10 20 50 Disp...
Страница 5: ...3 SPECIFICATIONS INDICATOR MEASUREMENTS 5 5 8 10 6 10...
Страница 6: ...me use for 10 12 hours to prevent low voltage resulted from self leakage of batter Once charged the battery should last for 45 hours To keep the battery in best condition fully discharge the battery e...
Страница 7: ...5 OP 907A LED OP 907B LCD...
Страница 8: ...On Off button to enter and exit calibration 2 Hold for 1 second to print The scale is at zero The scale is stable Gross Shows you are in Gross weight mode includes tare default mode Net Shows you are...
Страница 9: ...to tare out the weight Unit Selection To switch between measuring units kg lb oz press the UNITS key Tare Function The Tare function is used when you only wish to see the current change in weight not...
Страница 10: ...ll flash on the display followed by the total The total will display by flashing between 2 sets of numbers There are 8 digits in total the display will flash 4 at a time the first 4 on the left and th...
Страница 11: ...display the average weight sampled from 3 or 5 seconds Set in the C12 Parameter Add animal to scale and press HOLD Indicator screen will show LOC for 3 or 5 seconds then display the average weight fr...
Страница 12: ...n appears The C3 channel adjusts the divisions on the scale A value of 1 selected and C2 set to 1 the scale will read in 0 1 lb increments 13 Press SET to set the value The display will now show C04 1...
Страница 13: ...he scale will count down from 10 to 0 Once 0 has been reached the display will show CALEnd 27 Press SET to continue The display will now show C07 28 Press ON OFF to save and exit the setup menu 29 The...
Страница 14: ...labeled under each indicator button 3 Press the SET key to enter edit the parameter setting 4 Press the ON OFF key to save and exit settings at any time Table 1 Calibration Parameter Settings Function...
Страница 15: ...ght 3 Auto hold Automatically holds data when stable 4 Average hold for animal weighing averages the weight from a sample of 3 or 5 seconds Set in parameter C12 Hold Time C12 If you chose C11 4 this s...
Страница 16: ...1 second 2 2 seconds 3 3 seconds Overload Range C24 00 turn off overload range 01 99d overload range setting d division Negative Display C25 0 9d 10 10 max capacity 20 20 max capacity 50 50 max capac...
Страница 17: ...the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the set calibrated range Adds a tolerance...
Страница 18: ...wire load cell connections are both okay Please contact us directly if you have other special needs for your application Quick Disconnect as shown below QUICK DISCONNECT CONNECTION DIAGRAM DB9 SERIAL...
Страница 19: ...t Communication Format is done using ASCII as shown below CR LF S 1 S 2 S 3 Data S 4 S1 weight status ST standstill US not standstill OL overload S2 weight mode GS gross mode NT net mode S3 weight of...
Страница 20: ...e correct weight within the defined range ERR2 During calibration the weight is below the minimum required weight The calibration weight minimum is 10 of the max capacity set in C04 Recommended to use...