User Manual

 

7ISW152 Spectrometer

 

 

 

 

 

 

Optics Focus Instruments Co., Ltd.

 

Содержание 7ISW152

Страница 1: ...0 10 User Manual 7ISW152 Spectrometer Optics Focus Instruments Co Ltd ...

Страница 2: ...ts 1 Features 2 2 Description 2 3 Specifications 3 4 Specifications of Optional Gratings 3 5 CCD Specifications 4 6 Optical Design 5 7 Slits Adjustment 6 8 Usage 6 9 Others 7 10 Utility Software 8 11 Accessories 9 ...

Страница 3: ...mber are separated to reduce stray light and pollution to optical components 2 Description This series of Spectrometers is a high performance economical and user friendly spectrometer an ideal instrument for research and OEM applications This series of Spectrometers uses an asymmetrical in plane Czerny Turner optical configuration The optical configuration is designed to ensure high resolution and...

Страница 4: ... Wavelength Selection Method Motorized Output Ports 1 with CCD detector Slits Micrometer Slit Slits Height 5mm Slits Width 10µm 3mm Communication Interfaces RS232 and USB Size 298mmx200mmx185mm Center Height of Light Path 140mm 5mm Weight 10kg 4 Specifications of Optional Gratings Grating Model Linear Dispersion nm mm Accuracy nm Repeatability nm Resolution nm Theoretical Spectral Range nm Mechani...

Страница 5: ...0 1000 660 2000 OG600 1250 830 2200 OG300 500 21 6 1 0 0 4 1 6 330 1000 0 4400 OG300 1250 830 2500 OG300 1800 1200 3600 OG300 3000 2000 4400 OG150 4000 43 2 2 0 0 8 3 2 2600 8000 0 8800 OG66 DB 97 4 6 1 8 7 3 2500 25000 0 20000 OG50 12000 130 6 2 4 9 6 8000 24000 0 26400 5 CCD Specifications Manufacturer TOSHIBA Pixel Size 8μm 200μm Pixel Number 3648 SNR 60dB A D 16Bit Exposing Time 1ms 100ms PC I...

Страница 6: ...and changes the off axis angle to correct coma improves symmetry of spectral lines and improves resolution Eliminating secondary dispersion is designed to restrain stray light The F 4 5 spectrometer is optimized to provide excellent stray light rejection while minimizing aberrations 7ISW152 spectrometer optical design ...

Страница 7: ...per revolution 8 Usage Cable Connection Connect 9 pin connector of spectrometer to the RS232 serial port of computer Or connect the spectrometer to computer by USB line If you have bought our 7IFW6 filter wheel please connect 15 pin connector of spectrometer to the 9 pin connector of 7IFW6 filter wheel Connect the CCD to computer by USB line After connecting all the cables firmly turn on the power...

Страница 8: ...are and select the actual RS232 port the communication will be successful You can use software to operate the instrument to automatically change the wavelength scan spectrum switch gratings and change filters if you have bought 7IFW6 filter wheel from us Please refer to the software manual The spectrometer should be power on when using CCD The wavelength range of CCD need to be set by the spectrom...

Страница 9: ...gas so you can keep filling it during the whole experiment to keep the nitrogen concentration and positive pressure in the instrument The excess gas will leak off through gaps 10 Utility Software The utility software is included at no extra cost with all models to control both the monochromator Spectrometer and filter wheel accessory The utility software provided with the monochromator Spectromete...

Страница 10: ...9 10 Utility Software Screens 11 Accessories 1 x 220V line cord or 110V line cord 1 x RS232 cable 1 x USB cable 1 x USB cable for CCD 2 x 3A fuse 1 x Nitrogen connector ...