6212A Manual Version 14.56.docx
Page 53 of 58
Lasercheck will measure different signals and calculate different roughness for these two surfaces with identical
Ra. The surface roughness pattern (machining operation) and resulting scatter pattern therefore must be an
integral part of any Lasercheck calibration.
Base Calibration
Lasercheck is shipped calibrated to traceable standards for measurement of directional ground and sanded
surfaces. Lasercheck can be calibrated for other surface finishing processes using an Excel spreadsheet
calibration tool provided to generate a machining / surface pattern specific calibration file.
Overview of Calibration Process
Calibration of the Lasercheck gage involves customer testing known surfaces with the Lasercheck.
Lasercheck factory engineers use a custom excel spreadsheet tool to create a calibration file using those
customer test values. Software in the Lasercheck instrument uses the new calibration file when it performs a
measurement to provide users with roughness values calibrated to their finishing process.
Graph of directional ground surfaces from Lasercheck Excel Calibration Tool
Once a calibration is performed, Lasercheck never requires re-calibration for a specific surface finishing process;
calibration never changes.