Opti-Sciences, Inc.
CCM-200 plus operator's manual
20
"Light Leak!"
Fix:
This message indicates that there is too much light present during the test zero phase. This may
be due to worn gaskets or non-flat samples. Return to Dealer for repair.
Calibration errors:
"Clean Sample Chamber!"
Fix:
This message occurs during a calibration if the amount of light through the chamber is below a
certain level. This may be due to dirt on the diffusers or debris in the sample area. It may also be
caused by a mechanical or electrical problem. Clean chamber and re-try first then contact
service.
"Source/Detector Fail!"
Fix:
This message indicates that the unit has detected a fault with one of its sensing components. It
should be referred back to the factory of inspection.
"Chamber Light Leak!"
Fix:
This will occur with worn gaskets or possibly a defective detector. Contact service for new
gaskets or other instructions.