  / humidity and hand-held temperature measuring device CO

2

-TEMP(-RH)-HMG

Oppermann Regelgeräte GmbH

  |  Im Spitzhau 1  |  70771 Leinfelden-Echterdingen, Germany 

Phone +49 711 72723560  |  Fax +49 711 7280527  | [email protected] | www.oprg.de

Troubleshooting

Can’t power on

Press 

 for more than 0.3 seconds and try again. Check 

whether batteries are in good contact and correct polarity, or 
the adaptor is well plugged.

Fixed readings

Check whether data hold function was activated. (HOLD icon at 
the left top)

Slow response

Check whether the air flow channels on the rear were blocked.

Error messages

E01:   CO

2

 sensor damaged. 

E02: The value is under range. 
E03: The value is over range.
E04:  The original data error results in this error (DP, WB)
E07:  Too low voltage to measure CO

2

. Replace batteries or use 

an adaptor.

E17: Retry CO

2

 calibration.

E31:  Temperature sensor damaged. 
E34: Humidity sensor damaged.