InTemp™ CX Temperature Logger (CX40x) Manual
19996-C
InTemp CX400 series loggers measure temperature in indoor monitoring applications. Designed to
meet World Health Organization (WHO) and Centers for Disease Control and Prevention (CDC)
guidelines, this logger is ideal for clinical applications, such as vaccine storage and pharmaceutical
manufacturing. This Bluetooth® Low Energy-enabled logger is designed for wireless
communication with a mobile device. Using the InTemp app, you can easily configure the logger
with one of four preset configurations designed for ambient storage, clinical refrigerator, freezer,
or refrigerator monitoring or you can set up a custom configuration for other applications. You
can also quickly perform daily logger checks, download reports, and monitor tripped alarms. Use
the built-in LCD screen to check the current temperature, daily maximum or minimum
temperatures, logging status, battery use, and more. The CX402 model uses a 2- or 4-meter probe
and is available with a 5, 15, or 30 mL glycol bottle (bottle holder included). It also contains an
internal sensor for ambient temperature monitoring. The CX403 model is available with an
internal sensor only.
Specifications
External Temperature Probe
Range
-40° to 100°C (-40° to 212°F)
Accuracy
±0.5°C from -15° to 80°C (±0.9°F from 5° to 176°F)
±1.0°C from -40° to -15°C (±1.8°F from -40° to 5°F)
±0.7°C from 80° to 100°C (±1.3°F from 176° to 212°F)
Resolution
0.024°C at 25°C (0.04°F at 77°F)
Drift
<0.1°C (0.18°F) per year
Response Time
Probe only: 1 minute 30 seconds typical to 90% in air moving 1 m/s
With 5 mL bottle: 10 minutes typical to 90% in air moving 1 m/s
With 15 mL bottle: 20 minutes typical to 90% in air moving 1 m/s
With 30 mL bottle: 25 minutes typical to 90% in air moving 1 m/s
Cable Length
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
Probe Dimensions
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches)
long, 3.18 mm (0.125 inches) diameter
Ambient Temperature Sensor
Range
-30° to 70°C (-22° to 158°F)
Accuracy
±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F)
±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F)
Resolution
0.024°C at 25°C (0.04°F at 77°F)
Drift
<0.1°C (0.18°F) per year
Response Time
15 minutes typical to 90% in air moving 1 m/s
Logger
Radio Power
1 mW (0 dBm)
Transmission Range
Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard
Bluetooth Low Energy (Bluetooth Smart)
Logger Operating Range
-30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
Logging Rate
1 second to 18 hours
Time Accuracy
±1 minute per month at 25°C (77°F)
Battery Type
Two AAA 1.5 V alkaline or lithium batteries, user replaceable
Battery Life
1 year, typical with logging interval of 1 minute. Faster logging intervals,
remaining connected with the InTemp app, excessive report generation,
numerous audible alarms, and paging all impact battery life.
Memory
128 KB (84,650 measurements, maximum)
Full Memory Download
Time
Approximately 60 seconds; may take longer the farther the device is
from the logger
InTemp CX Temperature
Logger
Models:
•
CX402-T205, 2 meter probe
and 5 mL glycol bottle
•
CX402-T215, 2 meter probe
and 15 mL glycol bottle
•
CX402-T230, 2 meter probe
and 30 mL glycol bottle
•
CX402-T405, 4 meter probe
and 5 mL glycol bottle
•
CX402-T415, 4 meter probe
and 15 mL glycol bottle
•
CX402-T430, 4 meter probe
and 30 mL glycol bottle
•
CX402-T2M, 2 meter probe
•
CX402-T4M, 4 meter probe
•
CX403
Included Items:
•
Double-sided tape (for
models with glycol bottles)
•
Two AAA 1.5 V alkaline
batteries
•
Battery door and screw
•
NIST Certificate of
Calibration
Required Items:
•
InTemp app
•
Device with iOS 8.3–9.2 or
Android™ 4.4, 5.0, or 5.1