background image

 

1-800-LOGGERS (564-4377) • 508-759-9500 

www.onsetcomp.com/support/contact 

© 2015–2017 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOmobile are registered 
trademarks of Onset Computer Corporation. App Store is a service mark of Apple Inc. Google Play is a trademark of 
Google Inc. Bluetooth is a registered trademark of Bluetooth SIG, Inc. All other trademarks are the property of their 
respective companies. 

This product has been manufactured by Onset Computer Corporation and in compliance with Onset’s ISO 9001:2008 
Quality Management System. 

Patent #: 8,860,569 

19388-F MAN-QSG-MX2001 

 

Tap a row in the In Range list to connect to the logger. 

 

8

Once connected, tap Configure to set up the logger. 

 

Set the Reference Water Level and Water Density in the Configure screen. 

a.

 

Tap Reference Water Level or Water Density and enter the current water level as the reference level in meters or feet. 
Water levels that are below the reference point should be entered as negative values. The reference water level 
entered will be associated with a data point taken upon tapping Start (step 11) to determine the correction factor to be 
used for water level data. 

b.

 

Select the water density for the deployment or enter your own value under Manual Input. 

c.

 

Tap Done to return to the Configure screen. 

10 

Scroll down in the Configure screen to select the measurement parameters to be logged. Differential pressure must be enabled 
to calculate water level. Temperature must be enabled to log absolute pressure. Temperature is also required if you selected the 
water density option “Fresh Water adjusted for Temp.” 

 

11 

Select any other logger settings as desired and then tap Start in the upper right corner of the Configure screen. The logger will 
begin logging data based on the settings you selected. 

 

12 

To offload data from the logger to your device, connect to it and tap Readout (shown in example in step 8). The data from the 
logger is saved to your device. 

Tap the Data Files icon and then tap the mini-graph to view a larger version of the graph or to share the file. 

 

For detailed specifications and information about this logger, refer to the complete product manual. Go to 
www.onsetcomp.com/manual/mx2001 or scan the code at left.

 

 

 

 