18
CONTROL SOLUTION TESTING
When the confirmation window is full, the meter will count down from 5 to 1.
Read your result
Your control solution result will then appear on the display, along with the date, time, unit of
measure, and the words CONTROL TEST.
Control solution results can be seen when reviewing past results, but are not included in
result averages.
Check if the result is in range
Compare the result displayed on the meter to the OneTouch® Vita™ Control Solution range printed
on the test strip vial
.
If your control solution result falls outside the expected range, repeat the test with a new test strip.
NARROW CHANNEL
TOP EDGE
Apply drop here
Confirmation window full
11/OCT/08 14:07
mmol/L
7
.
7
CONTROL TEST
(Example)
CAUTION:
If the words
CONTROL TEST
do not appear on the screen, repeat the test
with a new test strip. If the problem persists, contact Customer Care on 0800 121 200 (UK) or
1800 535 676 (Ireland).
(Example range)
OneTouch
®
Vita
™
5.7–7.6 mmol/L
CAUTION:
• If you continue to get control solution test results that fall outside the range printed on the test strip vial,
Do not
use the meter,
the test strips, or the control solution. Contact Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland).
• The control solution range printed on the test strip vial is for OneTouch® Vita™ Control Solution only. It is not a recommended
range for your blood glucose level.
Control solution should not be
applied on top of the test strip.
Out-of-range results may be due to:
• not following the instructions detailed in steps 1–5,
• use of control solution that is contaminated, expired, or past its discard date,
• use of a test strip that is contaminated, damaged, expired, or past its discard date,
• a problem with the meter, or
• control solution test was performed outside the correct temperature range (20–25°C).