102
Detailed information about your system
8
System Accuracy
Diabetes experts have suggested that glucose meters
should agree within 0.8 mmol/L of a laboratory method
when the glucose concentration is lower than 4.2 mmol/L,
and within 20% of a laboratory method when the glucose
concentration is 4.2 mmol/L or higher. However, experts
have suggested greater accuracy of Blood Glucose
Monitoring Systems is desirable. The OneTouch® Verio™IQ
Blood Glucose Monitoring System has demonstrated
accuracy within 0.67 mmol/L of the laboratory method for
glucose concentrations lower than 4.4 mmol/L, and within
15% of the laboratory method for glucose concentrations
4.4 mmol/L or higher. Samples from 100 patients were
tested using both the OneTouch® Verio™IQ System and the
YSI 2300 Glucose Analyzer laboratory instrument.