103
Detailed information about your system
9
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch Verio Flex™
System compares well with a laboratory method.
# of Subjects
# of Tests
Slope
Intercept
(mmol/L)
100
600
1.01
-0.02
95% CI
Slope
95% CI
Intercept
(mmol/L)
Std. Error
(S
y.x
)
(mmol/L)
R
2
1.00 to 1.01
-0.12 to 0.08
0.64
0.99