background image

11

TESTING YOUR BLOOD GLUCOSE

Applying blood and reading results

Once you have a blood sample and your meter shows the APPLY BLOOD screen, you are ready to obtain a blood glucose result. If your meter 
does not show the APPLY BLOOD screen, remove the unused test strip and re-start the test process. See 

Getting a blood sample from the fingertip

 

in Section 3. 

 Prepare to apply the sample

Keeping your finger extended and steady, move the meter and test strip toward the blood drop.

Do Not

 apply blood on the top of the test strip.

Do Not

 hold the meter and test strip underneath the blood drop. This may cause blood to run into the test port and 

damage the meter.

 Apply the sample

Line up the test strip with the blood drop so that the narrow channel on the edge of the 
test strip is almost touching the edge of the blood drop.

Gently touch the channel to the edge of the blood drop.

Be careful not to push the test strip against your fingertip or the test strip may not fill 
completely.

•  

Do Not

 smear or scrape the drop of blood with the test strip.

•  

Do Not

 apply more blood to the test strip after you have moved the drop of blood away.

•  

Do Not

 move the test strip in the meter during a test.

 Wait for the confirmation window to fill completely

The blood drop will be drawn into the narrow channel and the confirmation window should fill completely.

When the confirmation window is full, this means you have applied enough blood. Now you can move the  
test strip away from the blood drop and wait for the meter to count down from 5 to 1.

 CAUTION:

 You may get an ERROR 5 message or an inaccurate result if the blood sample does not fill the confirmation window completely. 

See 

Troubleshooting

 in Section 8. Discard the test strip and re-start the test process.

Fingertip

Narrow channel

Full

Not full

Confirmation  

window full

Содержание ULTRA2

Страница 1: ...AW 06647702A Rev date 11 2010 USER GUIDE Blood Glucose Monitoring System Replaces Owner s Booklet...

Страница 2: ...used with OneTouch Ultra Test Strips No test strip supplier other than LifeScan is authorised to grant such a licence The accuracy of results generated with LifeScan meters using test strips manufactu...

Страница 3: ...neTouch Ultra 2 System is intended for use outside the body invitro diagnostic use by people with diabetes at home and by health care professionals in a clinical setting as an aid to monitoring the ef...

Страница 4: ...s If any of these items are missing from your kit contact LifeScan Customer Care on Freephone 0800 121 200 UK or 1800 535 676 Ireland AVAILABLE SEPARATELY f Clear Cap g OneTouch Ultra Control Solution...

Страница 5: ...sert into test port METER Display mmol L is the pre set unit of measurement Up and down buttons Select or change information Indicates selection available using up or down button Test port Insert test...

Страница 6: ...r is already on press and hold for two seconds to turn the backlight on or off Turning your meter off There are several ways to turn your meter off Press and hold for five seconds Your meter will turn...

Страница 7: ...AGE screen When using the meter for the first time or after changing the meter battery you will automatically start in the LANGUAGE screen In other cases from the MAIN MENU press or to select SET UP a...

Страница 8: ...press or to select AM or PM To confirm your selection press 8 Confirm your settings The choice YES will be highlighted at the bottom of the screen If your settings are correct press to confirm and sa...

Страница 9: ...ent when you get a blood glucose result Also you will only be able to review an average for all results stored in the meter You will not need to select a type of result average to review See Section 5...

Страница 10: ...t If a flashing appears instead of a code number such as when you are first using the meter follow the instructions in step 3 to change to a numerical code 3 Match the code on the meter with the code...

Страница 11: ...Do Not return the used test strip to the vial after performing a test Do Not re use a test strip that had blood or control solution applied to it Test strips are for single use only Write the discard...

Страница 12: ...eep enough to get a sufficient blood drop for testing LANCINGDEVICE CAUTION To reduce the chance of infection Never share a lancet or a lancing device with anyone Always use a new sterile lancet each...

Страница 13: ...ourmeter in Section 2 When the APPLY BLOOD screen appears on the display you can apply your blood sample 7 Lance your finger Hold the lancing device firmly against the side of your finger Press the re...

Страница 14: ...ing the edge of the blood drop Gently touch the channel to the edge of the blood drop Be careful not to push the test strip against your fingertip or the test strip may not fill completely Do Not smea...

Страница 15: ...cethelancetprotectivecoveronahardsurfaceandpushthelancettip intothecover Removethelancetandplaceitinacontainerforsharpobjects Replacethelancingdevicecap Disposing of the used lancet and test strip It...

Страница 16: ...oblem with your blood glucose control Obtain and follow instructions from your health care professional immediately Repeated unexpected glucose results If you continue to get unexpected results check...

Страница 17: ...s to indicate an insulin reaction Palm Forearm CAUTION Before testing make sure to wash the puncture site with soap and water Getting a blood sample Getting a blood sample from your forearm or palm is...

Страница 18: ...o Not smear the blood sample If the drop of blood runs or spreads due to contact with hair or a line in your palm Do Not use the sample Try lancing again in a smoother area Choose a different puncture...

Страница 19: ...ff It needs to be turned on before you can use it See Turningtheflags commentsfeatureoffor on in Section 2 for instructions Add or change a meal flag If the flags comments feature is turned on the up...

Страница 20: ...appropriate comment The available choices are NO COMMENT STRESS NOT ENOUGH FOOD ILLNESS TOO MUCH FOOD FEEL HYPO MILD EXERCISE MENSES period HARD EXERCISE VACATION MEDICATION OTHER Highlight NO COMMEN...

Страница 21: ...agged BEFORE MEAL and if the result is flagged AFTER MEAL The meter stores a maximum of 500 blood glucose or control solution test results When the memory is full the oldest result is dropped as the n...

Страница 22: ...ou will not be able to perform a test when the meter is in communication mode WARNING To avoid a possible shock Do Not insert a test strip when the meter is connected to a computer with the OneTouch I...

Страница 23: ...press to go to the APPLY BLOOD screen When you do not make a change after five seconds the display will advance to the APPLY BLOOD screen Mark the test as a control solution test Press to change APPLY...

Страница 24: ...s in range Compare the result displayed on the meter to the control solution range printed on the test strip vial Each vial of test strips may have a different control solution range If the results yo...

Страница 25: ...up on the battery ribbon The white ribbon is for the meter battery and the black ribbon is for the backlight battery Insert the new battery With the side facing up toward you place the battery in the...

Страница 26: ...vial for instructions on determining the discard date Cleaning your meter To clean your meter wipe the outside with a soft cloth dampened with water and mild detergent Do Not use alcohol or another s...

Страница 27: ...ld below 6 C to work correctly Wait a few minutes and insert a new test strip If you do not get another TEMPERATURE ERROR message the meter is now within the operating range There is a problem with th...

Страница 28: ...nough power for a minimum of 100 more tests Test results will still be accurate but replace the battery as soon as possible Oneofthefollowingmayapply You may have high glucose and have tested in an en...

Страница 29: ...t i e greater than 10 mmol L In this situation repeat the test in a warmer environment with a new test strip as soon as possible For accuracy and precision data and for important information on limita...

Страница 30: ...icient for n tests Consult instructions for use Recycle 2 Do not re use Separate collection of batteries Guarantee LifeScan guarantees that the OneTouch Ultra 2 Meter will be free of defects in materi...

Отзывы: