12
WL: 500. 0nm
Number : 3
OK
Cancel
WL: 500. 0nm
I nser t 1# St andar d.
I nput 1# Conc: 00000
OK
Cancel
Load the Stored Curve
Press
▲
,
▲
to choose “Load Curve”, press
▲
,
▲
to choose the
curve, press
SELECT
(left) to confirm.
Cr eat e Cur ve
Load Cur ve
Cl ear Cur ve
OK
Cancel
C=1. 000xA+0. 2 500. 0 <
C=0. 02xA+0. 32 470. 0
OK
Cancel
C=KxA+B
WL.
Press
SELECT
(left) to enter the test mode after building or loading
standard curve.
Step 3.
Enter into Measuring Interface
Press
SELECT
(left) to enter into the Quantitation Measuring Interface.
r =0. 996, C=2. 267+1. 211
- 0. 100
0. 900
4. 000
0. 000
Test
Cancel
+
+
+
WL: 500. 0nm 0. 000A
No. WL Abs.
Conc.
Test
Cancel
Step 4.
Blank
Put the Reference in the light path, press
ZERO
to do blank.
Step 5.
Measure Samples
Put the sample to be tested in the light path, press
SELECT
(left) to