background image

EVBUM2292/D

www.onsemi.com

2

Figure 1. Thermal Grease

When installing the sensor be sure to notice the pins are
keyed so it can only be installed one way. 
After the sensor is installed you will need to install the
lens mount. All lens mounts are designed to counteract the
force from the cold finger spring which will try to push the
sensor out of the socket. 
The KAE

01093 imager board has a more complex lens

mount because it is part of a passive cooling system. Make
sure to use all 8 of the M2 screws you remove when you
disassemble the lens mount from the imager board. When
you reinstall them tighten each a little at a time so you are
always applying even pressure against the front of the
sensor.

Plug in Power and Communication Cables

With the Evaluation Kit powered off, insert the USB cable
into your PC. Insert the power plug into the receptacle on
the Imager/FPGA board. Turn on the Power. Use
DeviceManager to verify “Truesense USB3” seen by the
system.

TE Cooler Controller (Recommended)

Connect the controller to the TE cooler interface
connector using the supplied pigtail (wires, contacts and
5 pin housing). The internally cooled kits have the
interface connector on the imager board. The externally
cooled kit has a short cable going to the thermistor and
TEC. Pinout for the cooler and thermistor is shown below:

Pin Number

Connection

1

Thermistor red

2

Thermistor black

3

No Connection

4

TE Cooler Red (pos)

5

TE Cooler Black (neg)

The external TE Cooler is a 2 stage TE Technologies

TE

2

127

127

1.3.  The thermistor is a 15K TE

Technology MP

2444 15. Use of the TE Technology

TC

48

20 controller is recommended. Recommend

cooling to +5

_

C for the KAE

0215X externally cooled kit

and to 0

_

C for the KAE

01093, KAE

0215X, KAE

0447X

and KAE

0815X internally cooled kits. If frosting or

condensation occur raise the operating temperature of the
TE controller 2

3 degrees. TC

48

2 controller has a DC

power supply input range of 0 to 50 V. Whatever input you
supply is the output it will use to drive the TEC. 

Do not

violate the voltage levels below:

Kit:

Voltage:

Current:

KAE

01093 Internal TEC

11.2 V

1.1 A

KAE

0215X internal TEC

11.2 V

1.1 A

KAE

0447X and

KAE

0815X internal TEC

14 V

1.4 A

TE CONTROLLER POWER SUPPLY:
Use an adjustable power supply. DO NOT run the 11.2 V
TEC devices with a 12 V supply. This reduces efficiency of
TEC and sends too much current through it. Recommend
adjustable power brick such as Ellway 4

24 V, 2.5 A, 2.1 x

5.5 connector, part number HGJL

1509T.

VERY IMPORTANT: 
Do not drive the TEC cooler with the TEC controller unless
the fan in the cooling tower is running. You can damage to
TEC cooler.

Run SensorStudio

Shortcut

To properly configure the shortcut icon for SensorStudio,
right click, choose properties and then select the
compatibility tab. Set the OS for your OS and enable “run
as admin”. The latter is needed if you install SensorStudio
in the standard Program Files area which is owned by
admin.

Launch Program

 

Double click the SensorStudio desktop icon to launch the
software.

Load Plugin

 

The plugin can be loaded by clicking the G3 button in the
toolbar. 

The KAE

01093 sensor will use the KAE1093 plugin.

The KAE

0215X uncooled sensor will use the KAE

plugin. The KAE

0215X cooled sensor will use the

Отзывы: