OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
33
MAIN UNIT WATER PRESSURE
The water pressure regulator is used to keep the water pressure of the instrument in the table constant. The
regulator is calibrated during the testing phase at O.M.S. and calibration can be changed exclusively for
technical reasons. This operation must be performed by an authorised technician.
Note:
The regulator has a filter that must be checked periodically and replaced, if necessary, as
described in paragraph MAIN UNIT WATER FILTERS.
WATER FLOW TO CUP AND SPITTOON
Inside the spittoon unit, a block is installed including four solenoid valves that are mounted on a fitting, which
contains the water filter. Each solenoid valve features a tap for controlling the water flow. The tap is adjusted
using a screwdriver to reduce the water flow when turned clockwise, and increase it when turned
anticlockwise:
A.
Free;
B.
Regulation of cold water to the cup;
C.
Regulation of water to the spittoon;
D.
Free;