DUKE EASY S.T.01/3AW Rev.0 01/2017
12
Outside each box there is a flag that indicates the order confirmation numbera and the model.
Note: the packaging of an eventual seat O.M.S. is separated.
ASSEMBLY
1.
Remove the equipment from its packaging;
2.
Place the chair on the previously made settings using the installation plan and adjust any “play” by
tightening or unscrewing the adjustment feet as necessary, secure the chair to the floor using the
two supplied dowels;
3. If an operating mounting lamp and electrical connection;
4. verify (and possibly perform) ground cables to the screw marked with the ground symbol;
5. to connect the vacuum pump (see the wiring diagram) in the same manner as described in the next
bullet for power cables (securing leads and traction).
6. remove the cover at the base of the dental chair, unscrewing the screws D
to maintain water and air connections (see section
developed), aspiration, exhaust and finally the power grid (Terminal); the power cables on the terminal
board must be general, with nylon ties, insulating specially drilled and placed under the Terminal, in
particular it must be ensured that, in the event of failure of the fasteners (clamps), the protective
conductor is not subject to traction as long as network cables are still connected to terminal
3.
Fit the accessories (instruments), if any, on the instrument and dental assistant tables.
N.B.: Refer to the wiring and installation diagrams provided for all electrical connections.