The breathable air shall meet the following standards of purity. If not specified otherwise the
contaminants shall be kept to a minimum, but in any event shall not exceed the permissible
exposure level.
The mineral oil content shall be such that the air is without odor of oil.
NOTE:
The odor threshold is in the region of 0.3mg/m³.
Compressed breathing air shall have a dew point sufficiently low to prevent condensation
and freezing. Where the apparatus is used and stored at a known temperature the pressure
dew point shall be at least 5 °C below the likely lowest temperature. Where national or state
regulations exist, they shall be observed.
Where the conditions of usage and storage of any compressed air supply is not known the
pressure dew point shall not exceed −11 °C.
Table 2-
Water content of high pressure breathing air
Nominal maximum supply pressure
bar
Maximum water content of air at
atmospheric pressure and 20°C
mg m
-3
40 to 200
≤ 50
> 200
≤ 35
The water content of the air supplied by the compressor for filling 200 bar or 300 bar
cylinders should not exceed 25
mg m
-3
.
Table 3-
Water content for supplied breathing air up to 40 bar
Nominal maximum supply pressure
bar
Maximum water content of air at
atmospheric pressure and 20°C
mg m
-3
5
290
10
160
15
110
20
80
25
65
30
55
40
50