Part Names and Functions
58
ZW-8000/7000/5000
User's Manual
Calibration ROM
This ROM is associated with the sensor on a one-to-one basis, and operates connected to the Sensor
Controller.
• Use with the Calibration ROM always connected. If the Calibration ROM is not connected, the error message
“SYSERR” is shown in the main digital and “NO.ROM” is shown in the sub digital, where measurement is not
possible.
Errors Common to All Communication States p.280
• If the Calibration ROM is lost or broken, as a provisional measure, the measurement can be resumed using the data
of the previously read Calibration ROM. For the operating procedure, refer to page
.
5-3 Recovering Calibration ROM data p.195
No.
Names
Functions
1
Serial number
Serial number.
Only a Sensor Head with the same serial number is available.
1
Important
Содержание ZW-7000 Series
Страница 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Страница 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Страница 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Страница 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 289: ...9 APPENDICES APPENDICES 9 1 Specifications and External Dimensions 288 9 2 Laser safety 323 9 3 Firmware Update 324 ...
Страница 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Страница 335: ......