Error Messages
280
ZW-8000/7000/5000
User's Manual
8-1 Error Messages
Errors Common to All Communication States
These errors occur in common regardless of communication state. When these errors occur, the ALARM
output of parallel I/O turns ON, and “SYSERR” and error code are displayed on the main and sub-displays,
respectively.
The acquired error number of EI command is showed in the number in parentheses.
If an error code other than the one listed above is displayed, the displacement sensor may be broken. Please
contact an OMRON branch or sales office.
*1:
As a provisional measure, the measurement can be resumed using the data of the previously read Calibration ROM.
<Operation method>
With error code NO.ROM displayed on the sub-display, hold down the Mode switching key, and then when [OK/CAN] is displayed, press
the ZERO/SET key.
<Cautions>
- When using this method, always check the serial number of the previously read Calibration ROM in "controller information," then check
that it matches the Sensor Head side serial number. Measurement will not be correct unless they match.
- When restarting the main unit, perform the same operations again.
This operation is disabled for a displacement sensor into which no Calibration ROM has ever been inserted and started up.
Error Code
Name
Description
Cause
Remedy
BRK.ROM
(02)
Linearity correction
data error
The linearity correction
data is corrupted.
Calibration ROM damage Check to make sure that the
Calibration ROM is correctly
inserted. If correctly inserted, the
Calibration ROM or displacement
sensor may be broken.
Please contact an OMRON branch
or sales office. (*1)
NO.ROM
(03)
Linearity correction
data read error
Reading of the linearity
correction data was not
executed correctly.
Calibration ROM not
inserted
BRK.SYS
(07)
System setting error
The system settings
saved to the Sensor
Controller are corrupt.
The displacement sensor
power was turned OFF
during saving/loading of
system settings.
After holding down the Mode
switching key, press the ZERO/SET
key to clear the system settings and
the bank data, and then resume the
starting process.
BRK.BNK
(08)
Bank data error
The bank data saved to
the Sensor Controller is
corrupt.
The displacement sensor
power was turned OFF
during saving/loading of
bank data.
BRK.TIM
(12)
Energization time
error
The energization time
data saved to the Sensor
Controller is corrupt.
The energization time is
corrupt due to a hardware
malfunction.
After holding down the Mode
switching key, press the ZERO/SET
key to resume. (The energization
time function becomes disabled.)
OVER.IP
(23)
IP address overlap
error
The same address as that
of the displacement
sensor exists on the
network.
Note: If a network hub is
not used, this error may
not appear.
The settings of IP
addresses are invalid.
Change the IP address so that the
IP address does not overlap with
devices on the network.
MIS.IP
(25)
Ethernet
communication
parameter error
An invalid IP address is
set.
Invalid IP address setting
Change to the correct IP address.
MIS.TYP
(60)
Type mismatch
The Sensor Head and
Sensor Controller types
do not match.
The Sensor Head and the
Sensor Controller are not
the correct pair. Both must
be either the ZW-8000
series or ZW-7000 series
or ZW-5000 series.
Replace with the correct Calibration
ROM. After replacing the Calibration
ROM, turn ON themeasurement
sensor power again.
MIS.SET
(61)
Set model mismatch
For the set model type,
the Sensor Controller and
Calibration ROM are not
matched.
A Calibration ROM except
ZW-7000 series is
inserted.
Insert the Calibration ROM for ZW-
7000 series, and then retry to turn
ON the Displacement Sensor.
MIS.BNK
(81)
Bank data error
The opened bank data is
abnormal.
Occurs as a result of
software downgrade.
To recover from the error, press and
hold
↑
key or
↓
key to clear the bank
data.
Содержание ZW-7000 Series
Страница 22: ...20 ZW 8000 7000 5000 User s Manual Index 329 Revision History 332 ...
Страница 23: ...ZW 8000 7000 5000 User s Manual 1 21 ...
Страница 54: ...Basic Operations for Displaying Measurement Results 52 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 96: ...Saving a project 94 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 192: ...Saving The Settings 190 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 280: ...Connecting by No protocol Communications 278 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 281: ...8 Troubleshooting Troubleshooting 8 1 Error Messages 280 8 2 Troubleshooting 282 ...
Страница 288: ...Troubleshooting 286 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 289: ...9 APPENDICES APPENDICES 9 1 Specifications and External Dimensions 288 9 2 Laser safety 323 9 3 Firmware Update 324 ...
Страница 330: ...Firmware Update 328 ZW 8000 7000 5000 User s Manual MEMO ...
Страница 333: ...Index ZW 8000 7000 5000 User s Manual 331 9 APPENDICES ...
Страница 335: ......