6
APPENDICES
Error Messages and Corrective Actions
77
ZG2-DSU User’s Manual
Error Messages and Corrective Actions
The following shows error messages that are displayed on the LCD screen and their
corrective actions.
Error message
Display details
Cause
Remedy
SYSTEM ERROR
BANK DATA
Internal bank data is in error.
Hold down the
key for 3
seconds, and then hold down the
key for 3 seconds.
After initialization ends, turn the
power OFF then back ON again
to restore the unit.
SYSTEM ERROR
CONNECT
The Sensor Controller is not
connected.
Connect the Sensor Controller.
SYSTEM ERROR
MAIN COM
Internal error
Turn the unit OFF then back ON
again.
File not found
The specified bank file does not
exist on the Memory Card.
Change the bank file number to
an existing file name.
Setting failed
The specified setting value is out
of range.
Set within the allowable setting
range.
Data error
The specified bank file is in error. Select the correct bank file.
Transfer failed
Bank file communication error
Check the connection between
Sensor Controllers.
Dead Battery
The clock backup battery has
run down.
Please contact your OMRON
sales representative.
Disp range Error
The measurement results
exceed the displayed number of
digits.
Change the number of digits
past the decimal point setting.
NO MEDIA
The Memory Card is not
inserted.
Insert the Memory Card.
Insufficient
There is not enough remaining
space on the Memory Card.
•Replace the Memory Card.
•Delete unwanted files from the
Memory Card.
•Format the Memory Card.
Содержание ZG2-DSU
Страница 1: ...User s Manual Cat No Z289 E1 01 Smart Sensor Data Storage Unit ZG2 DSU ...
Страница 10: ...8 ZG2 DSU User s Manual MEMO ...
Страница 14: ...12 ZG2 DSU User s Manual MEMO ...
Страница 38: ...Loading Logging Data to an External Device 36 ZG2 DSU User s Manual MEMO ...
Страница 42: ...Transferring Bank Data to the Controller 40 ZG2 DSU User s Manual MEMO ...
Страница 52: ...Bank Functions 50 ZG2 DSU User s Manual MEMO ...
Страница 86: ...Updating the Firmware 84 ZG2 DSU User s Manual MEMO ...
Страница 89: ...6 APPENDICES INDEX 87 ZG2 DSU User s Manual MEMO ...