PRINTER UNIT:

Power Source:

AA

batteries

Battery Life:

Approximately 300 uses (current data)

Paper Usage:

Approximately 200 uses for 1 roll (current data)

Printer Weight:

Approximately 170 g (not including batteries)

Printer Dimensions:

Approximately 123 mm x 72 mm x 50 mm

NOTE:

These specifications are subject to change without notice.

*Optional accessories: large cuff (fits arms 32 to 38 cm circumference) and small cuff (fits

arms 17 to 22 cm circumference) are sold separately. 

SPECIFICATIONS

Model:

T9P

Display:

LCD Digital Display

Measurement Range:

Pressure: 0 to 299 mmHg

Pulse: 40 to 180/minute

Accuracy/Calibration:

Pressure: 

±

4 mmHg 

Pulse: 

±

5% of reading

Inflation:

Automatic by electric pump

Deflation:

Active electronic control valve

Rapid Pressure Release:

Active electronic control valve

Pressure Detection:

Capacitive pressure sensor

Measurement Method:

Oscillometric method

Pulse Wave Detection:

Capacitive pressure sensor

Power Source:

4 "AA" batteries or optional AC adapter 

Battery Life:

Approximately 300 uses 

Operating 

10

°

C to 40

°

C

Temperatures/Humidity:

30 to 85% RH maximum

Storage

20

°

C to 60

°

C

Temperatures/Humidity:

10 to 95% RH maximum

Main Unit Weight:

Approximately 380g not including batteries

Main Unit Dimensions:

Approximately 115 mm (l) x 177 mm (w) x 71 mm (h)

Cuff Dimensions:

Approximately 140 mm x 480 mm

Cuff Circumference:

Fits arm circumference 22 ~ 32 cm

Accessories:

Arm cuff, printer-unit, paper rolls (2), instruction manual

MAIN UNIT:

T9P new  (9)  8/1/02  3:06 PM  Page 23

Содержание T9P

Страница 1: ...Automatic Blood Pressure Monitor and Print Out with IntelliSense Model T9P INSTRUCTION MANUAL R T9P new 6 7 17 02 6 05 PM Page 1 ...

Страница 2: ...Date 8 How to Use the Optional AC adapter 9 Correct Usage 10 How to Apply Arm Cuff 10 11 How to Take a Reading 12 13 How to Use the Memory Function 14 Printing 15 How to Connect Printer Unit 15 How to Load Printer Paper 16 How to Use the Printer Unit 17 18 19 Troubleshooting 20 21 Care and Maintenance 22 Specifications 23 About Blood Pressure 24 T9P new 6 7 17 02 6 05 PM Page 2 ...

Страница 3: ...reading Arm discomfort and misreading due to improper cuff inflation are virtually eliminated Your new OMRON monitor uses the oscillometric method of blood pressure measurement This means the monitor detects your blood s movement through your brachial artery and converts the movements into a digital reading At the push of a button the monitor automatically fills the cuff with air starts deflation ...

Страница 4: ...d of time approximately three months or more take out the batteries Replace the worn batteries with new ones immediately Do not use worn and new batteries together Do not insert the batteries with their polarities in wrong direction Do not force to bend the arm cuff or the air tube excessively When removing the air tube pull at the connector Do not apply strong shock or drop the main unit Do not i...

Страница 5: ...tton AC Adapter Set Date Time Adjust Date Time Diastolic Blood Pressure Print All Data Print Current Data Stop Printing Printer Cover Air Plug Connects to Air Jack Main Unit Printer Unit Accessories AC Adapter Jack Printer Jack Cuff Jack Printer Cable Plug Power Operation Light Applicable arm cuff for Adults 22 to 32 cm not a standard accessory T9P new 6 7 17 02 6 05 PM Page 5 ...

Страница 6: ...e brachial artery 5 Pull the end of the cuff so that the entire cuff is evenly tightened around your arm and press the hook material firmly against the pile side of the cuff 6 Press the ON OFF button 7 After the Heart Symbol appears on the digital panel press the Start button and remain still until the measurement is complete 8 When measurement is complete the monitor displays your blood pressure ...

Страница 7: ...e polarities match the polarities of the battery compartment as indicated 3 Replace the battery cover Notes If the Low Battery Indicator appears on the display replace all four batteries Long life alkaline batteries are recommended Remove the batteries if the monitor will not be used for an extended period of time Main Unit Printer Unit T9P new 6 7 17 02 6 23 PM Page 7 ...

Страница 8: ...ear digits 2001 will flash 3 Press the ADJUST button to advance the digit s one at a time If you hold down the ADJUST button the digits will advance rapidly 4 Press SET button when desired number is on the display to lock setting 5 Repeat Step 3 4 for month and date 6 Repeat Step 3 4 for hour and minutes T9P new 6 7 17 02 6 05 PM Page 8 ...

Страница 9: ...ug into the AC Adapter jack on the back of the main unit 2 Plug in the AC adapter into the appropriate AC outlet To remove the AC adapter To remove the AC adapter disconnect the adapter plug from the AC outlet first and then disconnect the cord from the monitor s jack NOTE The monitor is designed not to draw power from the batteries when the AC adapter is in use 1 2 T9P new 6 7 17 02 6 05 PM Page ...

Страница 10: ... loop 3 Sit in a chair with your feet flat on the floor and place your arm on a table so that the cuff is at the same level as your heart 4 Put your arm through the cuff loop The bottom of the cuff should be approximately 1 2 cm above the elbow The Green Marker on the cuff should lie over the brachial artery on the inside of the arm Tube should run down center of arm approximately even with middle...

Страница 11: ...od contact with your skin You should be able to fit your index finger between the cuff and your arm easily so you can pull the cuff off and on 8 Relax your arm and turn your palm upward 9 Be sure there are no kinks in the air tubing You can also measure blood pressure on your right arm Wrap the arm cuff as shown in the Figure 1 Place the air tube below the elbow 2 Place the mark cloth inside the e...

Страница 12: ...rement is started As the cuff slowly deflates decreasing numbers appear on the display and the Heart Symbol flashes at every heartbeat In rare circumstances a higher inflation may be necessary In those cases the monitor reinflates the cuff up to 30 mmHg higher than initial inflation and restarts the measurement 4 When the measurement is complete the arm cuff completely deflates and your blood pres...

Страница 13: ...inflates 30 to 40 mmHg higher than your suspected systolic pressure NOTE Do not apply more pressure than necessary The monitor will not inflate above 300 mmHg If you want to stop the measurement push the ON OFF button The monitor will stop inflating and start deflating rapidly then the monitor will turn off Instructions for special conditions T9P new 6 7 17 02 6 05 PM Page 13 ...

Страница 14: ... the Heart Symbol is displayed When twenty eight 28 sets of readings are stored in memory the oldest set will be deleted to store a new set START Press these two buttons at the same time Date and Time of stored readings will be alternately displayed To delete stored readings press the button and the Start button simultaneously then all the readings will be deleted NOTE You cannot delete the stored...

Страница 15: ...CONNECT PRINTER UNIT 1 Remove printer cable plug from around printer unit 2 Connect printer cable plug cord to the main unit Arrow 8 on printer cable plug should face down to main unit T9P new 6 7 17 02 6 05 PM Page 15 ...

Страница 16: ... the printing paper into the paper feeding slot When the edge of the paper protrudes sufficiently release the PAPER FEED button 4 Close the plastic printer cover Make certain that the paper sticks out from the top when the cover is replaced 5 Thermal sensitive paper is supplied with the unit Use only thermal paper of similar size width approx 38mm and roll diameter not to exceed approx 26mm These ...

Страница 17: ... to be ON to operate printer unit Printer unit will automatically shut OFF when printing is completed If you need to stop printing during printing process press DATA STOP button 1 Press DATA STOP to print measurement of most recent reading The numeric print out will look like the following example 7 9 6 36 AM SYS 111mmHg DIA 67mmHg PULSE 77 min Month Date Systolic Reading Diastolic Reading Pulse R...

Страница 18: ...adings Month s The month or months will be printed Example If the blood pressure reading was stored between months March June then the printout will indicate 3 6 Date Only the day s date 1 31 will be printed Example If the blood pressure reading was stored on June 1st then the printout will indicate 1 under the D column Time Hour Only the hour 24 hour format will be printed Example if the blood pr...

Страница 19: ...8 readings 50 100 150 200 mmHg 3 6 Diastolic Reading Systolic Reading Average systolic reading of stored measurements Average diastolic reading of stored measurements Last reading stored Month s If reading is lower than 50 mmHg or higher than 200 mmHg or will appear Most recent reading T9P new 6 7 17 02 6 05 PM Page 19 ...

Страница 20: ... under How to Take a Reading If EE is displayed again take another measurement by pressing and holding the START button as shown on instruction page 13 under How to Take a Reading instructions for special conditions Your cuff maybe wrapped too tightly around your arm Turn the monitor off carefully read and repeat steps listed under How to Take a Reading Replace all four AA batteries with new batte...

Страница 21: ... install again Purchase new printer paper Printer Unit TROUBLESHOOTING Batteries are worn No readings stored Printer paper is not installed correctly Replace with new batteries Take measurement then print Install correctly Batteries are worn Power Operation light is blinking stopped printing Replace with new batteries Nothing is printed No printer paper comes out No printer paper installed Printer...

Страница 22: ...oisture and direct sunlight Folding the cuff and tubing tightly Inflating the monitor over 280 mmHg Disassembling the monitor Subjecting the monitor to strong shocks for example dropping the monitor on the floor Do not clean the monitor with volatile liquids THE MONITOR SHOULD BE CLEANED WITH A SOFT DRY CLOTH Do not clean cuff T9P new 6 7 17 02 6 05 PM Page 22 ...

Страница 23: ...ulse 5 of reading Inflation Automatic by electric pump Deflation Active electronic control valve Rapid Pressure Release Active electronic control valve Pressure Detection Capacitive pressure sensor Measurement Method Oscillometric method Pulse Wave Detection Capacitive pressure sensor Power Source 4 AA batteries or optional AC adapter Battery Life Approximately 300 uses Operating 10 C to 40 C Temp...

Страница 24: ...er difficult to record the fundamental blood pressure but it is important to get readings in an environment that is as close as possible to this Dinner Sleep Rest room Breakfast Lunch Blood pressure mmHg Systolic blood pressure Diastolic blood pressure Time Brachial blood pressure data Fluctuation within a Day Example 35 year old male 300 250 200 150 100 50 0 17 00 23 00 5 00 11 00 17 00 mmHg Acco...

Страница 25: ...Product of OMRON Corporation Japan OMRON Healthcare Singapore Pte Ltd 83 Clemenceau Avenue 11 01 UE Square Singapore 239920 T9P new 6 7 17 02 6 05 PM Page 25 ...

Страница 26: ...T9P new 6 7 17 02 6 05 PM Page 26 ...

Страница 27: ...T9P new 6 7 17 02 6 05 PM Page 27 ...

Страница 28: ...T9P new 6 7 17 02 6 05 PM Page 28 ...

Отзывы: