• Do not touch the main unit for other than necessary operation such as turning off the 

power while nebulizing.

• To avoid the medication residue on the face, be sure to wipe the face after removing the 

mask.

• Do not use a damaged nebulizer kit or mouthpiece.
• While using this device, make sure that no mobile phone or any other electrical devices 

that emit electromagnetic fields is within 30 cm. This may result in degradation of 
performance of the device.

(Risk of electrical shock)

 

• Always unplug the AC adapter from the power outlet after use and before cleaning.
• Plug the device into the appropriate voltage outlet. Do not overload power outlets or use 

extension cords.

• Always remove the AC adapter from the device after use and before cleaning.
• Do not misuse the power cord of AC adapter.
• Do not wind the power cord around the compressor or AC adapter.
• Changes or modifications not approved by OMRON HEALTHCARE will void the user 

warranty.

• Do not pull the power cord of AC adapter strongly.

(Cleaning and disinfecting)

Observe the rules below when cleaning or disinfecting parts. Failure to observe these rules 
may result in damage, inefficient nebulization or infection. For the instructions, refer to 
“Cleaning and disinfecting” section.
• Do not use a microwave oven, dish dryer or hair dryer to dry the device or the parts.
• Do not use an autoclave, EOG gas sterilizer or low temperature plasma sterilizer.
• When disinfecting parts by boiling, make sure that the container does not boil dry. 

Otherwise it may also result in a fire.

General Safety Precautions:

• Inspect the device and parts before using them each time, and check that there are no 

problems. In particular, be sure to check the following: 
- That the vaporizer head, nozzle, kids accessory or air tube are not damaged.
- That the nozzle is not blocked.
- That the compressor operates normally.

• When using this device, there will be some noise and vibration caused by the pump in the 

compressor. There will also be some noise caused by the emission of compressed air from 
the nebulizer kit. This is normal and does not indicate a malfunction. 

• Operate the device only as intended. Do not use the device for any other purpose.
• Do not use the device at temperatures greater than +40°C.
• Make sure that the air tube is securely attached to the compressor and nebulizer kit, and 

does not come loose. Twist the air tube plug slightly when inserting it into the air tube 
connectors to avoid the air tube disconnecting during use.

• The AC adapter is provided so that the device can be completely isolated from the power 

source. To completely isolate the device from the power source unplug the plug from the 
power source.

• Read and follow the “Correct Disposal of This Product” in the Technical Data Section 

(page 25) when disposing of the device and any used accessories or optional parts.

Keep these instructions for future reference.

Содержание NE-C801KD

Страница 1: ...emploi Gebrauchsanweisung Manuale di istruzioni Manual de instrucciones Gebruiksaanwijzing TM IM NE C801S KDE V 05 03 2021 1887051 1F Issue Date Date de publication Ausgabedatum Data di pubblicazione...

Страница 2: ...ibed medication Make sure the vaporizer head is securely attached to the inhalation top Put the inhalation top back and rotate it clockwise until they click in place page 12 2 Assembling kids accessor...

Страница 3: ...ge 13 5 Applying mask and starting inhalation Apply the mask securely Turn on the power to the main unit page 14 Insert the power plug of AC adapter into the power connector of the main unit Plug the...

Страница 4: ...essful treatment of asthma chronic bronchitis allergies and other respiratory disorders The compressor forces air to the nebulizer kit When the air enters the nebulizer kit it converts the prescribed...

Страница 5: ...t This product is intended for use in a medical facility such as Hospital clinic and doctor s office and in a room of general household Durable period Durable periods are as follows provided the produ...

Страница 6: ...osed to noxious fumes or volatile substances Do not use the device where it may be exposed to flammable gas Do not cover the compressor with a blanket towel or any other type of cover during use Make...

Страница 7: ...one Do not spill liquid or medication on the compressor Do not tilt the nebulizer kit at an angle of greater than 45 degrees in all directions or shake it while in use Do not use or store the device w...

Страница 8: ...an autoclave EOG gas sterilizer or low temperature plasma sterilizer When disinfecting parts by boiling make sure that the container does not boil dry Otherwise it may also result in a fire General S...

Страница 9: ...Compressor Main Unit Cap Inhalation Air Inlet Inhalation Top Vaporizer Head Medication Tank Air Tube Connector Nozzle Rear View Nebulizer Kit Nebulizer Kit Kids Accessory Smile Bear Nebulizer Kit Hol...

Страница 10: ...10 1 Know your unit Other included items Spare Air Filters 5 Mouthpiece Child Mask PVC Infant Mask PVC AC Adapter Storage Bag Instruction Manual Warranty Card...

Страница 11: ...he air filter is changed regularly Refer to Section 7 Changing the air filter page 20 2 1 Connecting AC adapter 1 Make sure that the power switch is turned off Check that the power switch is in the of...

Страница 12: ...on tank Caution The capacity of the medication tank is 2 to 7 ml Note Scales on the nebulizer kit are for reference only The scale on the outside of the tank is with no vaporizer head in place Please...

Страница 13: ...air tube 1 Twist the air tube plug slightly and push it firmly into the air tube connector on the compressor 2 Twist the air tube plug slightly and push it firmly into the air tube connector on the m...

Страница 14: ...r during use Caution Do not block the air filter cover Limit use to 20 minutes at a time and allow a 40 minute interval before using the device again When using the device the main unit may become hot...

Страница 15: ...ore the air tube while it contains condensation or moisture This could result in infection via bacteria If there is moisture or liquid remaining in the air tube be sure to follow the procedure below t...

Страница 16: ...detergent neutral detergent Rinse them thoroughly with clean hot tap water and allow to air dry in a clean place Vaporizer Head Wash it with running water I Non washable parts Compressor Air Tube PVC...

Страница 17: ...ory as a result of repeated washing and disinfecting Handling the Vaporizer Head The vaporizer head is an important part that is used to vaporize medication Be sure to observe the following precaution...

Страница 18: ...ecting ethanol Sodium hypochlorite o o o o o X X Milton 0 1 15min B Boiling o o o X X X X an example of commercially available disinfectant The concentration and residence time specified in the table...

Страница 19: ...r under direct sunlight Do not crease or fold the air tube Dispose of the device and any used accessories or optional parts according to applicable local regulations Never clean with benzene thinner o...

Страница 20: ...Notes Use only OMRON air filters designed for this device Do not operate without a filter There is no front or back orientation for the air filters Check that air filters are clean and free of dust b...

Страница 21: ...orizer head missing and not assembled correctly Attach the vaporizer head correctly Is the nebulizer kit assembled correctly Assemble the nebulizer kit correctly Is the nozzle blocked Make sure that t...

Страница 22: ...rt to the manufacturer and the competent authority of the Member State in which you are established about any serious incident that has occurred in relation to this device Product Category Nebulizers...

Страница 23: ...cation of connector polarity Fuse For indoor use only Alternating current Direct current Certified by T V Efficiency Level of power supply The company logo of AC adapter supplier The company logo of A...

Страница 24: ...pensions or high viscosity See medication supplier s data sheet for further details See web site of OMRON HEALTHCARE EUROPE to update technical information URL www omron healthcare com Particle Size M...

Страница 25: ...the product or its literature indicates that it should not be disposed of with other household wastes at the end of its working life To prevent possible harm to the environment or human health from un...

Страница 26: ...izer Kit N 01 Smile Bear Set Model NEB NKKD1 81E V V T Nebulizer Kit N 01 Happy Bunny Set Model NEB NKKD2 81E Mouthpiece Model NEB MP E Child Mask PVC Model NEB MSMP E Infant Mask PVC Model NEB MSSP E...

Страница 27: ...IETNAM CO LTD No 28 VSIP II Street 2 Vietnam Singapore Industrial Park II Binh Duong Industry Services Urban Complex Hoa Phu Ward Thu Dau Mot City Binh Duong Province Vietnam Subsidiaries OMRON HEALTH...

Страница 28: ...28...

Отзывы: