Removing condensation from the air tube

If there is moisture or liquid remaining in the air tube, be sure to follow the 

procedure below to remove any moisture from within the air tube.
1)  Make sure that the air tube is connected to the air connector on the 

compressor.

2)  Remove the air tube from the nebulizer kit.

3)  Turn on the compressor and pump air through the air tube to expel the 

moisture.

Changing the air filter

If the air filter has been used for more than 60 days, replace it with a new 

one.

1.

  Remove the air filter cover using a flat screwdriver 

as shown on the right. 

2.

  Change the air filter. 

Remove the old air filter and insert the new air filter.

3.

  Put the air filter cover back in place. 

Make sure it is attached properly.

Notes:

•  Use only OMRON air filters designed for this device. Do not operate without a filter.
•  Do not wash or clean the filter. If the air filter becomes wet, replace it. Damp air filters 

can cause blockages.

•  There is no front or back orientation for the air filters.
•  Check that air filters are clean and free of dust before inserting them.

Troubleshooting

In case of any of the below problems occur during use, first check that no 

other electrical device is within 30cm.

If the problem persists, please refer to the table below.

Problem

Cause

Remedy

Nothing happens when the 

power switch is turned on.

Is the power plug plugged into 

a power outlet?

Check that the plug is inserted 

in a power outlet. Unplug then 

reinsert the plug if necessary.

No nebulization or low 

nebulization rate, when the 

power is on.

Is there medication in the 

medication tank?

Add the correct amount of 

medication to the medication tank.

Is there too much/little 

medication in the medication 

tank?

Is the vaporizer head missing 

or not assembled correctly?

Attach the vaporizer head 

correctly.

Is the nebulizer kit assembled 

correctly?

Assemble the nebulizer kit 

correctly.

Is the nozzle blocked?

Make sure that the nozzle is free 

of blockages.

Is the nebulizer kit tilted at a 

sharp angle?

Make sure that the nebulizer kit is 

not tilted at an angle of more than 

30 degrees.

Is the air tube connected 

correctly?

Make sure that the air tube 

is correctly connected to the 

compressor and nebulizer kit.

Is the air tube folded or 

damaged?

Make sure that the air tube does 

not contain kinks.

Is the air tube blocked?

Make sure that the air tube is free 

of blockages.

Is the air filter dirty?

Replace the air filter with a new 

one.

The compressor is 

abnormally loud.

Is the air filter cover attached 

correctly?

Attach the air filter cover correctly.

The compressor is very hot.

Is the compressor covered?

Do not cover the compressor with 

any type of cover during use.

Are the ventilation slots 

blocked?

Do not block the ventilation slots.

Note: 

If the suggested remedy does not solve the problem, do not try to repair the device 

- the device are not user serviceable. Please contact an authorized OMRON 

representative for assistance.

Cleaning and disinfecting

  

Caution: Handling the Vaporizer Head

•  Always wash the vaporizer head after each use.
•  Do not use a brush or pin to clean the vaporizer head.
•  When disinfecting the parts by boiling, be sure to boil them in plenty of water.

•  Do not boil vaporizer head together with objects other than applicable nebulizer accessories.

Cleaning

Clean the parts after each use to remove residual medication. This will avoid 

inefficient nebulization or infection.

  Washable parts

• Nebulizer Kit, Mask (PVC), Mouthpiece, Nosepiece 

Wash them in warm water and mild detergent (neutral detergent). Rinse them 

thoroughly with clean hot tap water and allow to air dry in a clean place.

• Vaporizer Head

Wash it with running water.

  Non-washable parts

• Compressor, Air Tube (PVC)

Firstly, make sure that the power plug is unplugged from the power outlet. Wipe 

clean with a soft cloth moistened with water or mild detergent (neutral detergent).

• Air Filter

Do not wash or clean the air filter. If the air filter becomes wet, replace it. 

Damp air filters can cause blockages.

Disinfecting

Disinfect the parts once a week. If the parts are heavily stained, replace 

them with new ones.

To select a method for disinfection, refer to the table below.

o

: applicable 

X

: not applicable

Disinfecting 

method

Parts

Mouthpiece

Nosepiece

Nebulizer Kit

Adult Mask (PVC) Child Mask (PVC)

Air T

ube

(PVC, 100cm)

Air Filter

Materials

PP

PP

PP

Mask: PVC

(Phthalate free) Band: Rubber***

PVC

(Phthalate free)

Polyester

A

Alcohol

o

o

o

o

**

X

X

Disinfecting 

ethanol
Sodium 

hypochlorite

o

o

o

o

**

X

X

Milton*

(0.1%, 15min.)

Quaternary 

ammonium

o

o

o

o

**

X

X

Osvan*

(0.1%, 10min.)

Chlorhexidine

o

o

o

o

**

X

X

Hibitane*

(0.5%, 30min.)

Amphoteric 

Surfactant

o

o

o

o

**

X

X

Tego 51*

(0.2%, 15min.)

B

Boiling

o

o

o

X

X

X

*   an example of commercially available disinfectant. (The concentration and 

residence time specified in the table are under the conditions where the service 

life of the parts is tested with each disinfectant used as described in its instruction 

manual. Please note that the testing was not performed for the purpose of ensuring 

the effectiveness of the disinfectants. There is no intention of suggesting that such 

disinfectants be used. Conditions of use and ingredients of disinfectant vary for 

manufacturers. Please read the instruction manuals carefully before use and apply 

disinfectant to each part in an appropriate way. Please notice that service life of the 

parts may be shorter depending on conditions, environments and frequency of use.)

**   Remove the rubber band from mask before disinfecting.

***  The band not made with natural rubber-latex.

A.

  Use a commercially available disinfectant. Follow the instructions 

provided by the disinfectant manufacturer.

Note:

  Never clean with benzene, thinner or a flammable 

chemical.

B.

  Parts may be boiled between 15 to 20 minutes.

After boiling, carefully remove the parts, shake off excess 

water and allow to air dry in a clean environment.

How to use

 Warning:

Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before 

using them when as follows:

 -the first time after purchase

 -if the device has not been used for a long period of time

 -if more than one person uses the same device

For the instructions, refer to “Cleaning and disinfecting” section.

1. 

Make sure that the power switch is in the off (   ) 

position.

2. 

Plug the power plug into a power outlet.

Note

Do not place the device in a location where it is difficult to disconnect 

the power cord.

3. 

Remove the inhalation top from the medication tank.

1) Rotate the inhalation top anti-clockwise.
2) Lift the inhalation top out of the medication tank.

1

2

4. 

Add the correct amount of prescribed medication to the 
medication tank. 

5. 

Make sure that the vaporizer head is placed in the 
medication tank.

6. 

Put the inhalation top back onto the medication tank.

1) Lower the inhalation top on the medication tank.
2) Rotate the inhalation top clockwise.

1

2

7. 

Attach the mask, mouthpiece or nosepiece to the nebulizer 
kit tightly.

8. 

Attach the air tube. Twist the air tube plug slightly and 
push it firmly into the air tube connector.

9. 

Hold the nebulizer kit as indicated below.

Follow the instructions of your doctor or respiratory therapist.

 Caution:

Do not tilt the nebulizer kit at an angle of greater than 30 degrees in all 

directions. Medication may flow into the mouth or it may result in not 

nebulizing effectively. 

Right angle

10. 

 Turn the power switch to the on (   ) position. As the 

compressor starts, nebulization begins and aerosol is 

generated.

Inhale the medication.

11. 

 When treatment is complete, turn the power off and 

unplug the compressor from the power outlet.

Important information regarding Electro Magnetic Compatibility (EMC)

NE-C101 manufactured by OMRON HEALTHCARE Co., Ltd. conforms to  

EN60601-1-2:2015 Electro Magnetic Compatibility (EMC) standard.

Further documentation in accordance with this EMC standard is available at  

http://www.omronhealthcare-ap.com/emc-information.

Refer to the EMC information for NE-C101 on the web site.