background image

 WARNING

 

Failure to properly clean the MICROAir components will result in the risk of bacterial 

growth, serious illness from contamination, compromised medication delivery.

 

Clean the parts of this device before using this device for the first time; after this device 

has not been used for an extended period of time and when parts are contaminated.

 

Clean the parts of this device after each use.

 

Allow the parts to dry after cleaning before reassembling and store in a clean location so 

that this device will not become contaminated.

 

DO NOT leave the cleaning solution in the parts. Rinse the parts with clean tap water 

after cleaning.

 CAUTION

 

DO NOT push or poke at the mesh with items such as a cotton swab or pin.

 

Dry the parts by air-drying only. DO NOT use any equipment to dry the device parts.

 

To prevent damage to the device, DO NOT immerse the Main Unit in any liquid, DO NOT wash or 

rinse any of the parts under strong running water, and DO NOT touch the mesh with your hand or any 

object.

 

DO NOT use household bleach to clean the device or parts; doing so will cause rust and damage the 

Mesh Cap.

 

To prevent damage to the device, DO NOT clean the Main Unit using abrasive cleaners or any type 

of chemical.

 

NEVER clean with benzene or thinner.

Before Using the Device for the First Time

Before using the device for the first time, disassemble the parts and clean the mesh cap, 

medication container, mouthpiece, mask, mask adapter and vibrating element and its 

surrounding surface as shown below:

Mesh Cap

 

Wash the mesh cap in mild (neutral) detergent, then rinse off with distilled water. 

 

Remove excess water and allow to air dry in a clean place.  Some types of medications may remain 

on the mesh.

 

Make sure to clean the medication from the mesh. 

 

Do not wash under running water.

Model 

NE-U100

 Vibrating Mesh Nebulizer

How to Clean the Device

3609065-5A

Отзывы: