GP-HEM-7213-E(V)-01-11/2011
Part 8
Comparison Table
New model
Predecessor
Model
M3W (HEM 7213-E(V))
M6 (HEM 7211-E / E8)
Measurement
Method
Oscillometric
Same as new model
Measurement
range
Pressure: 0mmHg to 299mmHg
Pulse: 40 to 180/min.
Same as new model
Memory
100 measurements
with date and time for each user (A and B)
90 measurements
with date and time for one user
Accuracy
Pressure: ±3mmHg
Pulse: ±5% of display reading
Same as new model
Inflation
Fuzzy-logic controlled by electric pump
Same as new model
Deflation
Automatic pressure release valve
Same as new model
Power source
4 “AA” batteries 1.5V or AC/DC adapter
Same as new model
Protection
Against Electric
Shock
Internally powered ME equipment
(When using only the batteries)
= Class II ME equipment
(Optional AC adapter)
Same as new model
Operating
temperature/
Humidity
+10
C to +40
C
Maximum: 30% to 85% RH
Same as new model
Storage
temperature/
humidity
-20
C to +60
C
Maximum: 10% to 95% RH / 700-1060hPa
Same as new model
Conclusion of Comparison
The new model is a substantial equivalent of both of its predecessor in terms of Technology,
Functionality, Safety, Reliability, Performance and general application.