Indication
Error
Cause
Correction
Correct reading
Cuff pressure low
1
Wait for at least three
not possible
minutes to repeat
Movement
measurement.
Inflate the cuff
Very weak artery
40 mmHg higher
1
/
pulse
30 mmHg lower
2
than previous value.
Cuff pressure
Follow instructions,
too high
2
repeat measurement
Air pressure
Air pressure
Press air release button
in cuff
not released
Low voltage
Batteries weak/
Fit four new identical
exhausted
1.5V alkaline batteries
type AA LR6
Display empty
8
M1 Classic
English
6 Error: causes and corrections
• Use a soft, slightly moistened cloth to
clean the OMRON M1 Classic.
Caution
! Do not use petrol, thinners or
similar solvents!
• Use a soft, moistened cloth and soap
to clean the cuff.
Caution
! Do not wash the cuff!
Caution
! Do not use petrol, thinners or
similar solvents!
• Do not carry out repairs of any kind
yourself. If a defect occurs consult your
OMRON retail outlet or distributor as
mentioned on the packaging.
Calibration
• It is recommended to have the OMRON
M1 Classic inspected every two years
to ensure correct function and
accuracy. Consult your OMRON retail
outlet or OMRON distributor as
mentioned on the packaging.
Replace batteries
• Remove the battery out of the battery
7 Maintenance and spare parts