5.3 Setting the Function for the K3HB-C
5-23
5.3
Setting the Function for the K3HB-C
■
F1: Individual Inputs
The count in incremented on input A pulses and decremented on input
B pulses.
The count is incremented on the rising edge of input A and
decremented on the rising edge of input B. When both inputs A and B
turn ON at the same time, the count does not change.
The measurement value can be obtained using the following formula:
D = C
×
α
C
: Count
α
: Prescale value A or prescale value B
D
: Measurement value
Note:
If F1 (individual inputs) is used, both preset value A and preset
value B must be set.
•
Holding the Measurement Value
Turning ON the HOLD input temporarily stops the cumulative count
and holds the measurement value. The outputs are also held.
•
Resetting the Display Value
The display value can be zeroed by turning ON the RESET input or
press the MAX/MIN Key for 1 second or longer.
While the RESET input is ON, measurement is not performed, the
display shows “-----”, and all outputs are OFF.
•
Compensation Value Input
Use the compensation input to start measurement from the desired
value. The compensation value must be set in advance.
Initial setting level
Input A
0
1
2
3
2
1
1
Input B
HOLD
2
3
Count
Содержание K35-2
Страница 2: ......
Страница 24: ...Outline Section 1 Outline 1 8 ...
Страница 25: ...2 1 Preparations Section 2 Preparations 2 1 Mounting 2 2 2 2 Using I O 2 4 ...
Страница 54: ...Basic Application Methods Basic Application Methods Section 3 Basic Application Methods 3 18 ...
Страница 62: ...Initial Setup Section 4 Initial Setup 4 8 ...
Страница 148: ...Functions and Operations Section 5 Functions and Operations 5 86 ...
Страница 149: ...6 1 Troubleshooting Section 6 Troubleshooting 6 1 Error Displays 6 2 6 2 Countermeasures 6 3 ...
Страница 152: ...Troubleshooting Section 6 Troubleshooting 6 4 ...
Страница 185: ......