List of Error Codes
Error code
Explanation
How to correct
Inflation error
• When the press
u
re does not exceed 12
mmHg
w
ithin the set time after the start of
inflation.
• When the inflation does not reach the set
c
u
ff press
u
re
w
ithin the specified time after
the start of inflation.
• Is the connector of the air t
ub
e connecting the
c
u
ff and the main
u
nit loose? Connect it sec
u
rely,
then meas
u
re.
• Is the air t
ub
e connecting the c
u
ff and the main
u
nit
b
ent? Straighten it, then meas
u
re.
• If the c
u
ff is
w
rapped loosely, it
w
ill take longer to
inflate. Wrap the c
u
ff correctly and meas
u
re.
• Is the
b
ladder of c
u
ff leaking air? Replace the
b
ladder
w
ith ne
w
one (option).
Deflation error
• When the deflation speed is too fast d
u
ring
meas
u
rement.
• When the deflation speed is too slo
w
d
u
ring meas
u
rement.
• When the meas
u
rement does not finish
w
ithin the specified time after starting the
meas
u
rement.
O
v
erpress
u
re error
• The c
u
ff press
u
re exceeded 299 mmHg.
• Is the air t
ub
e connecting the c
u
ff and the main
u
nit
b
ent? Straighten it and meas
u
re.
Ins
u
fficient inflation error
• Blood press
u
re co
u
ld not
b
e meas
u
red d
u
e
to ins
u
fficient inflation after the
meas
u
rement ends.
• If the meas
u
rement is made
b
y setting the P-SET
to "AUTO", ask the patient not to mo
v
e his/her
arm or
b
ody d
u
ring the inflation. The press
u
re
v
al
u
e may
b
e set lo
w
er d
u
ring the "AUTO"
meas
u
rement
b
eca
u
se of the noise.
• Is the P-SET sec
u
rely set to "AUTO"? T
u
rn the
Kno
b
co
u
nterclock
w
ise as far as it goes
u
ntil yo
u
can hear a click so
u
nd.
• If the meas
u
rement is made
b
y
u
sing the
press
u
re
v
al
u
e setting method, set the
v
al
u
e to 30
to 40 mmHg higher.
Indetermina
b
le
b
lood press
u
re error
• Blood press
u
re co
u
ld not
b
e meas
u
red
e
v
en
w
hen the c
u
ff press
u
re reached the
specified press
u
re.
• If the c
u
ff is
w
rapped loosely, correct
meas
u
rement is not possi
b
le. Wrap the c
u
ff
correctly and meas
u
re.
P
u
lse
w
a
v
e small error
• P
u
lse
w
a
v
e
w
as too small.
• If the c
u
ff is
w
rapped loosely, correct
meas
u
rement is not possi
b
le. Wrap the c
u
ff
correctly and meas
u
re.
• If the c
u
ff is
w
rapped o
v
er a thick cloth or
b
y
rolling
u
p a jacket or a shirt, the p
u
lse
w
a
v
e
cannot
b
e meas
u
red. Do not
w
rap the c
u
ff o
v
er
clothings.
Meas
u
rement error
• Relationship
b
et
w
een systolic and diastolic
press
u
res
w
as a
b
normal.
• Check the patient if he/she has arrhythmia.
• Ask the patient not to mo
v
e his/her arm or
b
ody
d
u
ring the meas
u
rement.
P
u
lse rate error
• P
u
lse rate is not
w
ithin the range of 30 to
199
b
eats/min.
De
v
ice error
• This is an indication of tro
ub
le of the main
u
nit.
• Contact the store
w
here yo
u
p
u
rchased this mon-
itor or the nearest OMRO
N
dealer.