29

5.

Press the [START/STOP] button to turn your monitor off.

Note

• If the “Err” symbol appears, follow the instructions in 

the “OMRON connect US/CAN” app.

• Your monitor automatically turns off after 2 minutes.

• Be aware that OMRON will not be responsible for the 

loss of data and/or information in the app.

• “OMRON connect US/CAN” is the only app that we 

recommend to use with your monitor to transfer data correctly.

4.2

Taking a Measurement

Refer to sub-section 3.1 to start a measurement. Open the app on your 

smart device and follow the instructions.

Note

• Make sure that your monitor is within 16 ft. (5 m) of your smart device.

Transferring Data Automatically/Manually

• When Bluetooth is enabled, the readings will be transferred to your smart 

device within 1 hour after taking a measurement.

• To transfer the readings manually, press the 

button.

Содержание HEM-7151T-Z

Страница 1: ...INSTRUCTION MANUAL Silver Upper Arm Blood Pressure Monitor Model BP5250 ENGLISH ...

Страница 2: ...onitor without a Smart Device 25 4 Using Your Monitor with a Smart Device 28 5 Using Memory Function 31 6 Other Settings 33 7 Error Messages and Troubleshooting 35 8 Maintenance 40 9 Optional Accessories 42 10 Specifications 44 11 FCC ISED Statement 47 12 Limited Warranty 49 13 Guidance and Manufacturer s Declaration 51 ...

Страница 3: ...s monitor READ and UNDERSTAND all of these instructions If you do not understand these instructions or have any questions contact 1 800 634 4350 before attempting to use this monitor For specific information about your own blood pressure consult with your physician Intended Use This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population ...

Страница 4: ...Glossary For symbol information visit OmronHealthcare com symbols glossary Contraindications This monitor is contraindicated for use in ambulatory environments This monitor is contraindicated for use on aircraft ...

Страница 5: ...reat high blood pressure DO NOT use this monitor on an injured arm or an arm under medical treatment DO NOT apply the arm cuff on your arm while on an intravenous drip or blood transfusion DO NOT use this monitor in areas containing high frequency HF surgical equipment magnetic resonance imaging MRI equipment computerized tomography CT scanners This may result in incorrect operation of the monitor...

Страница 6: ... AC adapter if this monitor or the AC adapter cable is damaged If this monitor or the cable is damaged turn off the power and unplug the AC adapter immediately Plug the AC adapter into the appropriate voltage outlet DO NOT use in a multi outlet plug NEVER plug in or unplug the AC adapter from the electric outlet with wet hands DO NOT disassemble or attempt to repair the AC adapter Battery Handling...

Страница 7: ...mble or attempt to repair this monitor or other components This may cause an inaccurate reading DO NOT use in a location where there is moisture or a risk of water splashing this monitor This may damage this monitor DO NOT use this monitor in a moving vehicle such as in a car DO NOT drop or subject this monitor to strong shocks or vibrations DO NOT use this monitor in places with high or low humid...

Страница 8: ...g and storage transport temperature refer to section 10 DO NOT use this monitor after the durable period has ended Refer to section 10 DO NOT crease the arm cuff or the air tube excessively DO NOT fold or kink the air tube while taking a measurement This may cause an injury by interrupting blood flow To unplug the air plug pull on the plastic air plug at the base of the tube not the tube itself ON...

Страница 9: ...dapter into the outlet When unplugging the AC adapter from the outlet be sure to safely pull from the AC adapter DO NOT pull from the AC adapter cable When handling the AC adapter cable DO NOT damage it DO NOT break it DO NOT tamper with it DO NOT forcibly bend or pull it DO NOT twist it DO NOT use it if it is gathered in a bundle DO NOT pinch it DO NOT place it under heavy objects Wipe any dust o...

Страница 10: ...move batteries if this monitor will not be used for a long period of time If battery fluid should get in your eyes immediately rinse with plenty of clean water Consult with your physician immediately If battery fluid should get on your skin wash your skin immediately with plenty of clean lukewarm water If irritation injury or pain persists consult with your physician DO NOT use batteries after the...

Страница 11: ...11 1 Know Your Monitor 1 1 Contents Product Includes Monitor Easy Wrap ComFit Cuff 4 AA batteries storage case instruction manual quick start guide ...

Страница 12: ...12 1 2 Monitor and Arm Cuff Front of the monitor Display Forward Backward buttons Connection button Date Time Memory button START STOP button a c b e d a b c d e ...

Страница 13: ...13 Left side back and bottom of the monitor Arm cuff Upper arm circumference 9 17 22 42 cm Air jack AC adapter jack for optional AC adapter Battery compartment Air plug Air tube g h f f g h i i j j ...

Страница 14: ...14 1 3 Display and Symbols B C D E F G H O A J I K L M N ...

Страница 15: ...ll Once you pair your monitor with your smart device transfer your readings immediately before the monitor deletes the oldest reading Up to 80 readings can be stored in the internal memory of your monitor Systolic blood pressure reading Diastolic blood pressure reading Deflation symbol Appears during cuff deflation Heartbeat symbol Flashes while taking a measurement Date Time display Bluetooth ON ...

Страница 16: ...is defined as a rhythm that is 25 less or 25 more than the average rhythm detected while your monitor is measuring blood pressure Irregular heartbeat symbol may be displayed when the monitor detects arm movements Refer to section 7 for more information If it continues to appear we recommend you to consult with and follow the directions of your physician Movement error symbol Appears with your read...

Страница 17: ...battery symbol Flashes when batteries are low Depleted battery symbol Appears when batteries are depleted Pulse display Memory number Pulse rate appears after the measurement When pressing the button the memory number will also appear for approximately 1 5 seconds before the pulse rate shows again on the display M N O ...

Страница 18: ...an BP Category Systolic BP Diastolic BP Normal 120 mmHg and 80 mmHg Elevated 120 129 mmHg and 80 mmHg Hypertension Stage 1 130 139 mmHg or 80 89 mmHg Stage 2 140 mmHg or 90 mmHg Individuals with Systolic BP and Diastolic BP in 2 categories should be designated to the higher BP category BP indicates blood pressure based on an average of 2 careful readings obtained on 2 occasions Source ACC AHA 2017...

Страница 19: ... time manually or refer to sub section 4 1 to pair your monitor to set the date and time automatically Note When the symbol flashes on the display it is recommended you replace batteries To replace batteries turn your monitor off and remove all batteries Then replace with 4 new batteries at the same time Replacing batteries will not delete previous readings The supplied batteries may have a shorte...

Страница 20: ...tly installed the year flashes on the display of your monitor 1 If nothing is displayed press the button to display the year If the year is already displayed skip this step 2 Press the or button to adjust the year Press and hold the button to quickly advance the years Press and hold the button to quickly go backwards through the years 3 Press the button to confirm the year then the month will flas...

Страница 21: ...ure Avoid taking measurements during stressful times Measurements should be taken in a quiet place It is important to take measurements at the same times each day Taking measurements in the morning and in the evening is recommended Remember to have a record of your blood pressure and pulse readings for your physician A single measurement does not provide an accurate indication of your true blood p...

Страница 22: ...sure can differ between the right arm and the left arm and the measured blood pressure readings can be different OMRON recommends to always use the same arm for measurement If the readings between both arms differ substantially check with your physician to determine as to which arm to use for your measurements Caution Remove tight fitting or thick clothing from your arm while taking a measurement ...

Страница 23: ...iddle finger 3 Make sure that the air tube is positioned on the inside of your arm and wrap the cuff securely so it can not move around your arm To take a measurement on your right arm When you take a measurement on your right arm the air tube should run along the side of your elbow along the bottom of your arm Be careful not to rest your arm on the air tube Bottom edge 0 5 inch 1 to 2 cm Bottom e...

Страница 24: ...d in a room with a comfortable temperature Place your arm on the table Sit in a comfortable chair with your back and arm supported Keep your feet flat and your legs uncrossed The arm cuff should be placed on your arm at the same level as your heart with the arm resting comfortably on a table ...

Страница 25: ...top the measurement press the START STOP button once to deflate the arm cuff Caution DO NOT use this monitor with other medical electrical ME equipment simultaneously This may result in incorrect operation of the monitor and or cause an inaccurate reading Remain still and DO NOT talk while taking a measurement 1 Press the START STOP button All symbols appear on the display before starting the meas...

Страница 26: ...ed If either the systolic or the diastolic reading is high refer to sub section 1 3 the symbol appears 3 Press the START STOP button to turn your monitor off Note Your monitor automatically turns off after 2 minutes Wait 2 3 minutes between measurements The wait time allows the arteries to decompress and return to their pre measurement state You may need to increase the wait time depending on your...

Страница 27: ...isplay turns off The arm cuff automatically starts to inflate If your systolic pressure is more than 210 mmHg If your systolic pressure is more than 210 mmHg take a measurement manually After the arm cuff starts to inflate press and hold the START STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure Note Do not inflate above 299 mmHg Caution Inflating to...

Страница 28: ...tall the OMRON connect US CAN app onto your smart device If you already have the OMRON connect US CAN app and have created your account open the app and add your new monitor 3 Open the app and follow the pairing instructions shown on your smart device 4 Confirm that your monitor is connected successfully When your monitor is connected successfully to your smart device the OK symbol appears Review ...

Страница 29: ...CAN is the only app that we recommend to use with your monitor to transfer data correctly 4 2 Taking a Measurement Refer to sub section 3 1 to start a measurement Open the app on your smart device and follow the instructions Note Make sure that your monitor is within 16 ft 5 m of your smart device Transferring Data Automatically Manually When Bluetooth is enabled the readings will be transferred t...

Страница 30: ...ther information on potential restrictions refer to documentation on the Bluetooth usage by the FCC ISED Bluetooth is enabled as default 1 When your monitor is off press and hold the button for more than 10 seconds oFF flashes 3 times 2 Press the START STOP button to turn your monitor off Note To enable the Bluetooth press and hold the button for more than 2 seconds on appears on the display Your ...

Страница 31: ...in Memory 1 Press the button The Memory number appears for 1 5 seconds before the pulse rate is displayed The most recent reading stored in the memory will be numbered as 1 Note If the reading is high refer to sub section 1 3 the symbol appears If there are no readings stored in the memory the screen to the right is displayed If the memory is full new readings will replace the old readings startin...

Страница 32: ...t 2 or 3 readings taken within the most recent 10 minute timeframe Note Set your monitor to the correct date and time before taking a measurement to view the average readings 1 When your monitor is off press and hold the button for more than 2 seconds The symbol flashes Note If there are only 2 readings in the memory for that period the average will be based on these 2 readings If the average read...

Страница 33: ...or more than 3 seconds The display to the right flashes 3 While holding the button down again press and hold the START STOP button for more than 3 seconds The display stops flashing then all readings are deleted 4 Press the START STOP button to turn your monitor off Note All readings will be deleted You cannot partially delete the readings stored in the memory Your monitor automatically turns off ...

Страница 34: ...ing the button down again press and hold the START STOP button for more than 3 seconds The display stops flashing then your monitor is restored to the default settings 3 Press the START STOP button to turn your monitor off Note After the completion the year flashes on the display Refer to sub section 2 2 to set the date and time manually or refer to sub section 4 1 to pair your monitor to set the ...

Страница 35: ...n was pressed while the arm cuff is not applied Press the START STOP button again to turn the monitor off After inserting the air plug securely and applying the arm cuff correctly press the START STOP button Air plug is not completely plugged into the monitor Insert the air plug securely Refer to sub section 2 4 The arm cuff is not applied correctly Apply the arm cuff correctly then take another m...

Страница 36: ...ressure is above 210 mmHg and a measurement cannot be taken Inflate the arm cuff manually until the systolic pressure is 30 to 40 mmHg above your previous readings Refer to sub section 3 1 appears The arm cuff is inflated exceeding the maximum allowable pressure Do not touch the arm cuff and or bend the air tube while taking a measurement If inflating the arm cuff manually refer to the end of sub ...

Страница 37: ...cian appears does not flash during a measurement appears The monitor has malfunctioned Press the START STOP button again If Er still appears contact customer service Refer to section 12 appears The monitor cannot connect to a smart device or transmit data correctly Follow the instructions shown in the OMRON connect US CAN app If the Err symbol still appears after checking the app contact customer ...

Страница 38: ...n the OMRON connect US CAN app to transfer your readings flashes There are more than 64 readings in memory to be transferred Pair or transfer your readings to the OMRON connect US CAN app so you can keep them in memory in the app and this error symbol disappears The date and time is not set appears There are 80 readings in memory to be transferred flashes Batteries are low Replacing all 4 batterie...

Страница 39: ...nd or how you apply the arm cuff may affect your blood pressure Review sub sections 2 3 2 5 and section 3 Any other communication issue occurs Follow the instructions shown in the smart device or visit the Support section in the OMRON connect US CAN app for further help If the problem still persists contact customer service Refer to section 12 Any other problems occur Press the START STOP button t...

Страница 40: ...and other components in the storage case when not in use Store your monitor and other components in a clean safe location 1 Remove the arm cuff from the monitor Caution To unplug the air plug pull on the plastic air plug at the base of the tube not the tube itself 2 Gently fold the air tube into the arm cuff Note Do not bend or crease the air tube excessively Changes or modifications not approved ...

Страница 41: ...erse your monitor and arm cuff or other components in water Do not use gasoline thinners or similar solvents to clean your monitor and arm cuff or other components 8 4 Disposal Dispose of your monitor other components and optional accessories according to applicable local regulations Unlawful disposal may cause environmental pollution If your monitor and other components are wet In locations expos...

Страница 42: ...the air plug away The air plug can be applicable to the optional cuff For further information visit our website at OmronHealthcare com Easy Wrap ComFit Cuff AC Adapter Arm circumference 9 17 22 42 cm Model HEM FL31 B HEM FL31 BZ Model HEM ADPTW5 ...

Страница 43: ...ate and time each time you unplug and plug back the AC adapter The readings will not be deleted 1 Insert the AC adapter plug into the AC adapter jack on the rear side of your monitor 2 Plug the AC adapter into an electrical outlet To unplug the AC adapter unplug the AC adapter from the electrical outlet and then remove the AC adapter plug from the monitor Caution ONLY use the AC adapter arm cuff b...

Страница 44: ...flation Automatic pressure release valve Measurement method Oscillometric method Transmission method Bluetooth Low Energy Wireless communication Frequency range 2 4 GHz 2400 2483 5 MHz Modulation GFSK Effective radiated power 20 dBm Operation mode Continuous operation IP classification Monitor IP20 Optional AC adapter IP21 Rating DC6 V 4 0 W Power source 4 AA batteries 1 5 V or optional AC adapter...

Страница 45: ... Transport conditions 4 F to 140 F 20 C to 60 C 10 to 90 RH non condensing Weight Monitor Approximately 12 oz 340 g not including batteries Arm cuff Approximately 5 8 oz 163 g Dimensions Monitor Approximately 4 2 w 3 4 h 6 0 l 105 mm 85 mm 152 mm Arm cuff Approximately 5 7 20 9 air tube 29 5 145 mm 532 mm air tube 750 mm Cuff circumference applicable to the monitor 9 to 17 220 to 420 mm Memory Sto...

Страница 46: ...mmunication interference This product operates in an unlicensed ISM band at 2 4 GHz In the event this product is used near other wireless devices such as microwave and wireless LAN which operate on the same frequency band as this product there is a possibility that interference may occur If interference occurs stop the operation of the other devices or relocate this product away from other wireles...

Страница 47: ...erference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television recep...

Страница 48: ...ISED radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency RF Exposure Guidelines and RSS 102 of the ISED radio frequency RF Exposure rules as this equipment has very low levels of RF energy ...

Страница 49: ...ement is our only responsibility and your only remedy under the above warranty To obtain warranty service contact Customer Service by calling 1 800 634 4350 for the address of the inspection center and the return shipping and handling fee that may apply Enclose the original printed receipt Include a letter with your name address phone number and description of the specific problem Pack the product...

Страница 50: ...L DAMAGES SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU This warranty provides you with specific legal rights and you may have other rights that vary by jurisdiction Because of special local requirements some of the above limitations and exclusions may not apply to you FOR CUSTOMER SERVICE Visit our web site at For USA OmronHealthcare com For Canada OmronHealthcare ca Call toll free 1 ...

Страница 51: ...Documents in the Scope of IEC60601 1 2 2014 Important information regarding Electromagnetic Compatibility EMC BP5250 conforms to IEC60601 1 2 2014 Electromagnetic Compatibility EMC standard Further documentation in accordance with this EMC standard is available at OmronHealthcare com emc Refer to the EMC information for BP5250 on the website ...

Страница 52: ...d the Apple logo are trademarks of Apple Inc registered in the U S and other countries and regions App Store and App Store logo are service marks of Apple Inc Google Play and the Google Play logo are trademarks of Google LLC OMRON HEALTHCARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPAN Distributed by OMRON HEALTHCARE INC 1925 West Field Court Lake Forest IL 60045 U S A For USA OmronHea...

Отзывы: