
13
Important:
• Use only the OMRON glucose control solution HEA-CTR30 with HEA-232
meter and HEA-STP30 strips.
• Mark the newly opened bottle of control solution with the date opened; discard
any unfinished control solution over three months after opening.
• Store control solution at room temperature. Do not freeze.
• Shak e control solution before you use it. Please replace new control solution if
the control solution deteriorate or coagulate.
* Before running the test, place the meter, test strips and glucose control
solution at room temperature to reach the new temperature before use.
A.
Prepare for the HEA-232 and test strip
Insert the test strip into the meter. The meter will display the code number.
Make sure the code number on the display matches the code number on
the test strip vial. (Refer to Chapter 3 Fingertip blood measurement for more
information)
When the HEA-232 is ready to apply blood sample, press button twice to
enter control solution test mode. The “Blood” symbol will disappear in the
control solution test mode.
In the control solution test mode, the test result will not be recorded.
B. Prepare for the control solution
Before using the control solution, gently shake the control bottle to ensure it is mixed well.
Remove the control bottle cap and wipe the tip of the bottle with a tissue. Squeeze the
bottle until a tiny drop forms at the tip of the bottle and drop on a clean surface.
C. Run the control solution test
Touch the tip of the test strip to the drop of control solution. The control solution test result
will show on display after 5 seconds count down. The HEA-232 will automatically power
off after the test strip is removed. Discard the used test strip in accordance with local
regulations.
Finger tip blood test mode
Control solution test mode
OMRON blood glucose monitoring system HEA-232.indd 13
2012/6/13 下午 06:44:26