- The part where the cuff is wrapped is at a different height than the heart. (A
difference of 10 cm (4 inches) in height may cause a variation in the blood pressure
value of up to 7 or 8 mmHg.)
- Body movement or conversing during measurement.
- Cuff wrapped over thick clothing.
- Pressure on the arm due to a rolled up sleeve.
• In the case of a cuff for adults, the cuff should be wrapped to a tightness that allows
two fingers to be inserted in between the cuff and the arm.
• The accuracy of a flashing measurement value that is out of the measurement
range cannot be guaranteed. Always check the patient’s condition before deciding
what steps to take.
• Do not use the cuff if it is damaged or has holes.
• Only an OMRON GS CUFF2 can be used with this device. The use of any other
cuff may result in incorrect measurement.
Disposal
As there is a risk of environmental pollution, follow your applicable national and local
legal regulations regarding disposal or recycling of this equipment
.
The main
constituents of each part are listed in the table “Materials List”.
As there is a risk of infection, do not recycle patient attachments such as cuffs, but
dispose of them as instructed by your facility’s procedures and applicable
regulations.
Consult also the “
Correct Disposal of This Product (Waste Electrical & Electronic
Equipment
)” section of the Manual.
Transportation and storage conditions
Storage and transportation:
Temperature range: -20 to +60°C (-4 to 140°F)
Humidity range: 10 to 95%RH
Atmospheric pressure: 500 to 1060hPa
Warranty and lifetime
Warranty:
- blood pressure monitor: 3 years
- AC adapter: 1 years
- cuffs: this is considered a disposable item depending upon usage
Lifetime:
- blood pressure monitor: 5 years (only when appropriate inspection in performed)
- AC Adapter: 5 years
- tubes: no lifetime as these are considered disposable items