4. Troubleshooting and Maintenance
13
EN
The ambient temperature is not within the range
between 10°C and 40°C (50°F - 104°F).
Allow the thermometer to rest in a room for at least
30 minutes at room temperature: 10°C and 40°C
(50°F - 104°F).
Error 5-9, the system is not functioning properly.
Remove the battery, wait for 1 minute and repower
it. If the message reappears, please contact the
OMRON retail outlet or distributor for having the
device checked.
Ear measurement mode:
Temperature taken is higher than 42.2°C
(108.0°F).
Check the integrity of the probe cover and take a
new temperature measurement.
Ear measurement mode:
Temperature taken is lower than 34.0°C (93.2°F).
Check the integrity of the probe cover and take a
new temperature measurement.
Surface mode:
Temperature taken is higher than 80.0°C
(176.0°F).
Check the integrity of the probe cover and take a
new temperature measurement.
Error Display
Cause
Remedy
MC-521-E_A_M_main.book Page 13 Tuesday, March 26, 2013 2:41 PM
Содержание Gentle Temp 521
Страница 24: ...6 Some Useful Information 23 EN MC 521 E_A_M_main book Page 23 Tuesday March 26 2013 2 41 PM ...
Страница 48: ...6 Informations utiles 47 FR MC 521 E_A_M_main book Page 47 Tuesday March 26 2013 2 41 PM ...
Страница 72: ...6 Nützliche Hinweise 71 DE MC 521 E_A_M_main book Page 71 Tuesday March 26 2013 2 41 PM ...
Страница 96: ...6 Informazioni utili 95 IT MC 521 E_A_M_main book Page 95 Tuesday March 26 2013 2 41 PM ...
Страница 120: ...6 Información útil 119 ES MC 521 E_A_M_main book Page 119 Tuesday March 26 2013 2 41 PM ...
Страница 144: ...6 Nuttige informatie 143 NL MC 521 E_A_M_main book Page 143 Tuesday March 26 2013 2 41 PM ...
Страница 168: ...6 Полезная информация 167 RU MC 521 E_A_M_main book Page 167 Tuesday March 26 2013 2 41 PM ...
Страница 215: ......
Страница 216: ...HEM 6200 E_ en_TOC fm Page 14 Tuesday May 26 2009 2 00 PM ...