Controlling Operation and Outputting Data with TCP No-protocol Communications
FQ2-S4 User’s Manual
403
Con
necting
throu
gh
Eth
e
rnet
9
Setting the Data To Output Automatically after Measurements
You can set the data to output automatically after measurements. (You can set up to 32 data items.)
Data That Can Be Output
• Data Output
You can output up to 32 data items (data 0 to data 31).
The measurement data from inspection items that can be output and the calculation results from the
expression settings can be output. For data that can be output, refer to the
Measurement Data That Can Be
Used for External Outputs and Calculations
for each inspection item.
Assigning Detection Results to Output Data: p. 403
Assigning More Than One Detection Result to Output Data: p. 404
• Outputting Character Strings
You can output a character string that contains up to 1,024 characters for each of the inspection item that reads
a character string, such as the OCR inspection item. Also, when reading the character string fails, you can
output a specific character string that is set in advance.
Outputting Read Character Strings: p. 408
Assigning Inspection Results to Output Data
You can individually assign the parameters of the inspection items to output data (data 0 to data 31).
The following procedure shows how to assign the measured position X of [0. Search] to data 0 for a binary
output.
[In/Out]
−
[I/O setting]
−
[Output data setting]
−
[Link data output]
−
[Output data set]
1
Press [0. Data 0].
2
Press [Data setting].
3
Press [I0. Search].
4
Press [Position X X].
5
If the inspection item allows multi-point output,
press the number ([0] to [30]) of the inspection result
for which to output the data from the list of inspec-
tion results.
To register something to data 1 and higher, repeat
this process.
The settings will be enabled after you restart the Sen-
sor.
▲
Содержание FQ2-S4
Страница 1: ...User s Manual Smart Camera FQ2 S4 Cat No Z330 E1 01 ...
Страница 58: ...Setting Up Ethernet 56 FQ2 S4 User s Manual MEMO ...
Страница 214: ...Calculations and Judgements Using Inspection Item Data 212 FQ2 S4 User s Manual MEMO ...
Страница 234: ...Adjusting Judgement Parameters during Operation 232 FQ2 S4 User s Manual MEMO ...
Страница 446: ...Controlling Operation and Outputting Data with FINS TCP No protocol Commands 444 FQ2 S4 User s Manual MEMO ...
Страница 452: ...Controlling Operation and Outputting Data with RS 232C No protocol Communications 450 FQ2 S4 User s Manual MEMO ...
Страница 453: ...11 Troubleshooting Troubleshooting 11 1 Error Histories 452 11 2 Error Messages 454 11 3 Basic Troubleshooting 455 ...
Страница 458: ...Basic Troubleshooting 456 FQ2 S4 User s Manual MEMO ...
Страница 591: ...Index FQ2 S4 User s Manual 589 Index 12 ...
Страница 593: ......