18
4.Cleaning and disinfecting
o
: applicable X: not applicable
A.
Use a commercially available disinfectant. Follow the instructions provided by the
disinfectant manufacturer.
Note:
Never clean with benzene, thinner or a flammable chemical.
B.
Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake off excess water
and allow to air dry in a clean environment.
Disinfecting
method
Parts
M
out
hpiece
Nebulizer
K
it
A
ir
T
ube
(S
ilicone,
100cm
)
(opt
ional)
Infant
M
as
k
(P
V
C)
Child
M
ask
(P
VC)
K
ids Acces
sor
y
A
ir
T
ube
(P
VC,
100c
m)
A
ir
F
ilt
er
Materials
PP
PP
(V
apor
izer
H
ead:
P
C)
T
ube: Silicone
Connector
: P
P
M
ask:
P
VC
(P
htha
late f
re
e)
B
and:
Rubber
***
PP
, PV
C
(P
htha
late f
re
e)
PV
C
(P
htha
late f
re
e)
Polyes
te
r
A
Alcohol
o
o
o
o
**
X
X
X
Disinfecting
ethanol
Sodium
hypochlorite
o
o
o
o
**
o
X
X
Milton*
(0.1%, 15min.)
Quaternary
ammonium
o
o
o
o
**
o
X
X
Osvan*
(0.1%, 10min.)
Chlorhexidine
o
o
o
o
**
o
X
X
Hibitane*
(0.5%, 30min.)
Amphoteric
Surfactant
o
o
o
o
**
o
X
X
Tego 51*
(0.2%, 15min.)
B
Boiling
o
o
o
X
X
X
X
* an example of commercially available disinfectant. (The concentration and residence time
specified in the table are under the conditions where the service life of the parts is tested with
each disinfectant used as described in its instruction manual. Please note that the testing was
not performed for the purpose of ensuring the effectiveness of the disinfectants. There is no
intention of suggesting that such disinfectants be used. Conditions of use and ingredients of
disinfectant vary for manufacturers.
Please read the instruction manuals carefully before use and apply disinfectant to each part in
an appropriate way. Please notice that service life of the parts may be shorter depending on
conditions, environments and frequency of use.)
** Remove the rubber band from mask before disinfecting.
***The band is not made with natural rubber latex.
NE-C801S-KDE(V)_main.book Page 18 Friday, December 14, 2018 10:32 AM
Содержание CompAIR NE-C801KD
Страница 28: ...28 NE C801S KDE V _main book Page 28 Friday December 14 2018 10 32 AM ...
Страница 56: ...56 NE C801S KDE V _main book Page 56 Friday December 14 2018 10 32 AM ...
Страница 84: ...84 NE C801S KDE V _main book Page 84 Friday December 14 2018 10 32 AM ...
Страница 112: ...112 NE C801S KDE V _main book Page 112 Friday December 14 2018 10 32 AM ...
Страница 140: ...140 NE C801S KDE V _main book Page 140 Friday December 14 2018 10 32 AM ...
Страница 168: ...168 NE C801S KDE V _main book Page 168 Friday December 14 2018 10 32 AM ...
Страница 196: ...196 NE C801S KDE V _main book Page 196 Friday December 14 2018 10 32 AM ...
Страница 197: ...۲۸ AR ...