Device

 

Errors

 

Error

 

or

 

symbol

 

Cause

 

Resolution

 

Device

 

shows

 

“E2”

  

Cuff

 

is

 

under

inflated

 

or

 

over

inflated.

 

Single

 

Mode:

 

The

 

monitor

 

will

 

attempt

 

1

 

reading.

 

 

TruRead™

 

Mode:

 

The

 

monitor

 

will

 

attempt

 

up

 

to

 

5

 

times

  

 

After

 

the

 

automatic

 

attempts,

 

remove

 

the

 

arm

 

cuff.

  

Read,

 

“Taking

 

a

 

Measurement”

 

in

 

the

 

instruction

 

manual.

  

Take

 

another

 

measurement.

 

Device

 

shows

 

‘E1’

 

Air

 

plug

 

is

 

not

 

connected

 

or

 

air

 

is

 

leaking

 

from

 

the

 

arm

 

cuff.

 

Check

 

air

 

plug

 

connection.

  

Take

 

another

 

measurement

 

and

 

remain

 

still

 

until

 

the

 

measurement

 

is

 

complete.

 

Device

 

shows

 

“Battery

 

picture

 

with

 

a

 

cross

 

mark”

 

Batteries

 

are

 

worn

 

Replace

 

the

 

four

 

batteries.

 

Refer

 

to

 

“Battery

 

Installation”

 

in

 

the

 

instruction

 

manual.

 

No

 

Power

 

 

No

 

Display

 

appears

 

on

 

the

 

unit

 

Power

 

off

 

Check

 

that

 

the

 

device

 

is

 

turned

 

on.

  

 

Check

 

the

 

battery

 

installation

 

for

 

proper

 

placement

 

of

 

the

 

battery

 

polarities.

 

 

Replace

 

all

 

four

 

batteries

 

with

 

new

 

ones.

 

 

If

 

running

 

from

 

a

 

wall

 

outlet,

 

check

 

that

 

the

 

cord

 

is

 

attached

 

to

 

both

 

the

 

device

 

and

 

outlet

 

 