This manual is a guide for the use of the Omni International THq, Digital Tissue Homogenizer and accessories.

Data herein has been verified and validated. It is believed adequate for the intended use of the instrument. If the instrument or 

procedures are used for purposes over and above the capabilities specified herein, confirmation of the validity and suitability 

should be obtained, otherwise Omni International does not guarantee results and assumes no obligation or liability. This 

publication is not a license to operate under, or a recommendation to infringe upon, any process patents. Notes, cautions, and 

warnings within the text of this manual are used to emphasize important and critical instructions.

This Omni International product is warranted to be free from defects in material and workmanship for a period of THREE 

YEARS from the date of delivery. Omni International will repair or replace and return free of charge any part which is returned 

to its factory within said period, transportation prepaid by user, and which is found upon inspection to have been defective in 

materials or workmanship. For the first 90 days, both parts and service are without charge. For the balance of the period, parts 

will be provided but service will be charged at established labor rates. This warranty does not include normal wear from use; it 

does not apply to any instrument or parts which have been altered by anyone other than an employee of Omni International 

nor to any instrument which has been damaged through accident, negligence, failure to follow operating instructions, the 

use of electric currents or circuits other than those specified on the plate affixed to the instrument, misuse, or abuse. Omni 

International reserves the right to change, alter, modify, or improve any of its instruments without any obligation whatsoever 
to make corresponding changes to any instrument previously sold or shipped.

THE FORGOING OBLIGATION IS IN LIEU OF ALL OBLIGATIONS AND LIABILITIES INCLUDING NEGLIGENCE AND ALL WARRANTIES 

OF MERCHANTABILITY OR OTHERWISE, EXPRESSED OR IMPLIED IN FACT OR BY LAW, AND STATE OUR ENTIRE AND EXCLUSIVE 

LIABILITY AND BUYERS EXCLUSIVE REMEDY FOR ANY CLAIM OF DAMAGES IN CONNECTION WITH THE SALE OR FURNISHING 

OF GOODS OR PARTS, THEIR DESIGN, SUITABILITY FOR USE, INSTALLATION, OR OPERATION. OMNI INTERNATIONAL WILL 

IN NO EVENT BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES WHATSOEVER, AND THEIR LIABILITY UNDER NO 

CIRCUMSTANCES WILL EXCEED THE CONTRACT PRICE FOR THE GOODS FOR WHICH LIABILITY IS CLAIMED.

- READ ALL INSTRUCTIONS BEFORE USING.

The Omni THq has been engineered for economical functionality as well as safety; however, basic safety precautions 

and common sense must always be demonstrated when using any electrical product. DO NOT attempt to modify 

any part of the Omni THq. If you experience problems with or have questions about your Omni THq, contact your 

authorized dealer or call Omni International at 1-800-776-4431 or 1-770-421-0058.

- DO NOT allow the machine to be submerged in any liquid.

- DO NOT use in any setting other than an indoor laboratory.

- DO NOT plug power cord into an incorrect outlet. 

To reduce the risk of burns, electrocution, fire, or injury:

- Use this product only for its intended purpose as described in this  booklet. DO NOT use attachments not 

recommended by the manufacturer.

- DO NOT operate the product if it is damaged in any way.

- Keep this product away from heated surfaces.

WARNING: The tip of the saw tooth generator probe is sharp.

WARNING: Reduce the risk of unintentional starting; make sure that the power switch is in the OFF position prior to 

plugging in the unit.

WARNING: DO NOT process pathogenic material in an open container, since aerosols created during normal processing 

could be inhaled by the operator. Please call for assistance in processing pathogens or other material which require 

sealed enclosures.

WARNING: Keep all housings in place and in working order.

WARNING: Remove all tools from the generator probe before turning the motor on.

WARNING: DO NOT use the motor in a dangerous environment.

WARNING: Disconnect the motor before servicing, and when changing the generator probe.

WARNING: DO NOT modify the plug or cord that is provided. If the plug will not fit the outlet, have the proper outlet 

installed by a qualified electrician.

WARNING: Reduce the risk of unintentional starting; make sure the power switch is in the OFF position before 

plugging in the motor.

WARNING: Damaged or worn power cords should be repaired or replaced immediately by a qualified electrician.

WARNING: Improper connection of the equipment can result in a risk of electric shock. 

DECONTAMINATION

Should an instrument or component that has been used with radioactive or pathogenic material require factory or 

field service, comply with the following procedure to ensure the safety of service personnel:

Clean the parts to be serviced of all encrusted material and decontaminate them. There must be no radioactivity 

detectable by survey equipment. Obtain a Decontamination Certificate from Omni International. Complete the 

certificate and attach to the instrument or parts being returned. If no Decontamination Certificate is attached, and a 

potential radioactive or biological hazard is detected or suspected by Omni International, the equipment will not be 

serviced until proper decontamination and certification is complete. The sender will be contacted for instructions as to 

the disposition of the equipment. Disposition costs will be borne by the sender.

WARNING: It is a violation of federal law to transport biologically hazardous or radioactive materials without proper 

packaging, labeling, and appropriate warnings.

Should this product ever require service, please contact 

Omni International at 1-800-776-4431

Содержание THq

Страница 1: ...THq Digital Tissue Homogenizer User Manual OMNI THq...

Страница 2: ...ify any part of the Omni THq If you experience problems with or have questions about your Omni THq contact your authorized dealer or call Omni International at 1 800 776 4431 or 1 770 421 0058 DO NOT...

Страница 3: ...ssue Omni Tip 30007L 6 Hard Tissue Omni Tip 30007HT 6 Instruction Manual 03 225 1 Overview 1 2 3 4 5 6 7 8 1 Power cord port 2 Run Stop LED indicators 3 Power 4 GeneratorProbe coupling 5 Digital Reado...

Страница 4: ...the THq press the black button on the back of the opening of the post clamp assembly and lift the motor from the cradle 3 Slide the post clamp assembly down over the end of the post until the clamp is...

Страница 5: ...rator Probe into coupling 2 Quarter turn clockwise to lock 3 Ready to use always submerge probe in liquid before starting 1 4 Turn IMPORTANT INFORMATION The blue cap must be removed from the end of th...

Страница 6: ...B5 075 5 mm x 75 mm 6 pack 2 mL 5 mL B5 075 6 7 mm x 110 mm 25 mL 30 mL B7 110ST 7 mm x 110 mm 6 pack 25 mL 30 mL B7 110ST 6 10 mm x 110 mm 1 5 mL 100 mL B10 110ST 10 mm x 110 mm 6 pack 1 5 mL 100 mL...

Страница 7: ...connectors Check wall socket for power to the outlet Teflon bearing wears quickly Fluid level in the sample may be too low Immerse the probe further into the fluid Execessive splashing in the sample F...

Страница 8: ...PN 03 225 Rev C 935C Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...