SECTION 4 — TROUBLESHOOTING
Do not attempt to service the Omni TH
Q
in a manner other than those
discussed in this manual. For any issue that is unsuccessfully corrected
using this guide, please contact your authorized dealer or call Omni
International technical assistance at 1-800-776-4431 or 1-770-421-0058.
PROBLEM
CORRECTIVE ACTION
The Omni TH
Q
is plugged in and
turned on but is not functioning.
- Check power cord connectors
- Check wall socket for power
to the outlet
Teflon bearing wears quickly
- Fluid level may be too low
in the tube.
- Immerse the probe deeper
into the fluid.
Excessive splashing
in sample tube
- Fluid level too low for tube size
Motor does not spin.
Green light flashes.
Rotor is locked.
Remove the obstruction.