3.3 SEALED GENERATOR PROBE ASSEMBLIES
Sealed chamber generator probe assemblies are available for use with the Omni Mixer
and Macro but are not supplied with the instrument. They consist of chambers, many
of which are fluted to facilitate mixing by inhibiting swirling, gasket screw on cover
seals to inhibit aerosol release, and a generator assembly. Chamber materials include
stainless steel, glass, polycarbonate, and polypropylene. A full range of titanium
chamber assemblies is also available for applications requiring protection from leached
elemental components, or where additional strength is a factor. Please contact Omni
International at 1-800-776-4431 for complete details.
SECTION 3. SEALED CHAMBER ASSEMBLIES
3.4 INSTALLING SEALED CHAMBER ASSEMBLIES
All chamber assemblies are easily installed to the motor unit by following these steps:
1. Attach the chamber (with adapter if required) to the cover/generator probe
assembly. Method of attachment is a left-handed screw thread. Tighten the
assembly securely.
2. Attach the chamber assembly to the motor by screwing the chamber cover into
the coupling adapter.
3. If the chamber assembly does not thread fully into the coupling adapter, then
unscrew the chamber assembly from the coupling adapter. Slightly rotate the
rotor shaft collar on the generator probe. Then return to Step 2.
4. Before operating, make sure that the chamber assembly rests securely on the
tabletop.