26
Technical
c. Excessively high stimulation intensity - adjust appropriate level control.
The stimulator bleeps for 30 seconds then switches off:
a.
Low battery - replace battery if a non-rechargeable PP3 is used, recharge
battery if a rechargeable battery is used.
Specifications
Output amplitude:
Maximum 100mA with 1KOhm output load.
Stimulation frequency:
Programmable (20 or 40Hz).
Stimulation wave form:
Nominally square with passive charge
balancing.
Stimulation pulse width: 330µs maximum .
The Microstim2 is a Class IIa Medical Device within the definition of the European
Medical Devices Directive.
Expected Service Life and Shelf life
This is an indication of the life span of the parts (figures based upon daily use). Life
span will depend on usage.
4
Item
Expected service
life
Shelf life
Microstim 2V2
5 Years
N/A
Electrode socket
2 Years
N/A
Electrode lead
6-12 Months
N/A
Neurostimulation electrodes
3-4 Weeks
See expiry date
PP3 Battery (9V) (non-rechargeable)
6-12 Weeks
See expiry date