Test Procedures
34
Linearity Test (Calibration Verification Check)
The ‘Calibration Validation Check’, known as ‘Linearity Check’ on the OmiSmoke,
needs to be carried out every 7.5 days. The OmiSmoke prompts the user to carry out
a ‘Linearity Check’ when calibration is due, alternatively the user can carry out the
check on demand.
Note: If the ‘Linearity Check’ is carried out on demand by the user, OmiSmoke will
automatically calculate a new 7.5 default.
Once ‘Linearity Test’ is selected from the main menu, The OmiSmoke automatically
performs a tube warm up procedure followed by a contamination check and a smoke
bench reset before the ‘Linearity Check’ screen is displayed.
The message ‘Please insert filter’ appears on the ‘Linearity Check’ screen. Insert the
filter into the sample probe port in the back on the smokemeter, ensuring the filter is
inserted fully and in the correct orientation (see illustration).
Once the filter is inserted correctly, confirm by pressing the tick key.