User
Manual
Version
1.0
33
5.4
Starting
operation
Use
the
following
steps
to
setup
the
system
and
start
operation.
1.
Connect
the
external
power
supply
to
the
LightHUB®
ULTRA.
2.
Connect
the
external
power
supply
to
the
mains.
3.
Connect
the
loopback
connector
to
the
control
port
or
connect
a
customer
specific
circuit
to
close
the
key
switch
and
the
interlock
loop.
Alternatively
connect
the
optional
CDRH
kit.
4.
If
you
have
an
optional
heatsink,
make
sure
the
cable
is
plugged
into
the
Fan
Out
connector.
5.
Push
the
power
button
on
the
front
panel
of
the
LightHUB®
ULTRA.
Key
switch
is
off.
6.
The
fans
should
turn
on
and
the
blue
backlight
LED
of
the
power
button
starts
blinking.
7.
Turn
the
key
switch
to
ON
position.
Blue
backlight
LED
of
the
power
button
stops
blinking
if
system
operating
temperature
setpoint
is
reached.
8.
Connect
a
signal
to
the
safety
shutter
connector
or
use
a
bridge
connector
to
open
the
safety
shutter.
9.
Connect
a
logic
high
level
to
the
digital
modulation
input
and
an
appropriate
voltage
level
to
the
analogue
modulation
input
of
a
channel,
which
is
equipped
with
a
laser.
OR
Connect
the
system
to
a
personal
computer
via
USB
and
run
Omicron
Control
Center
software
to
change
operating
mode
of
the
lasers
to
CW
and
deactivate
digital
input.
WARNING!
Laser
emission
is
possible.
Assure
laser
safety
precautions.
The
system
should
emit
laser
light
now.
Содержание Light HUB Ultra
Страница 1: ......