Getting Started with Quick
3 - 13
Switch off on trigger
This option is available if the trigger condition is either binary or
overload. Else, "switch off on trigger" is preset and cannot be disabled.
"Switch off on trigger" is also locked and cannot be changed after
having pressed K
E E P
R
E S U L T S
.
1. Enabled
When the trigger event occurs, the
CPC 100
outputs are
switched off immediately (that is, in real-time).
The measurements are updated to show the values at
the time the trigger event occurred, and are then
"frozen", that is, not updated anymore from that moment
on. Quick turns to the "off" state.
At the field for the power output value setting, the data
entry mode is automatically disabled to prevent this
value from being changed. The focus, however,
remains on that field, enabling instant navigation with
the handwheel.
Pressing the menu key K
E E P
R
E S U L T S
saves the
measured values in the measurement table. Quick
returns to the "measuring" state, the measurement
continues in a new line of the measurement table.
2. Disabled
The
CPC 100
outputs remain active, that is, in "on"
state, after the trigger event occurred. If the trigger is
binary, the "BinIn" indicator reflects the occurrence of
the trigger event accordingly.
The focus remains in data entry mode on the field for the
power output value setting, allowing an immediate
change of the output value.
The measurements continue after the trigger event.
As soon as the output value is changed, the display of
the most recent delay time is removed.
5
Содержание CPC 100
Страница 12: ...CPC 100 V1 41 x...
Страница 28: ...CPC 100 V1 41 1 16...
Страница 90: ...CPC 100 V1 41 3 14...
Страница 194: ...CPC 100 V1 41 6 30...
Страница 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Страница 258: ...CPC 100 V1 41 9 4...
Страница 264: ...CPC 100 V1 41 10 6...
Страница 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Страница 284: ...CPC 100 V1 41 12 12...
Страница 312: ...CPC 100 V1 41 14 12...
Страница 316: ...CPC 100 V1 41 15 4...
Страница 350: ...CPC 100 V1 41 16 34...
Страница 370: ...CPC 100 V1 41 17 20 7 Finish the assembly by screwing the backnut to the housing and by tightening them with wrenches...
Страница 372: ...OMICRON Contact Addresses 22...