Resistance
7 - 3
Resistance - µ
Ω
Measurement
Use the
Resistance
test card to measure test objects with very low resistance,
such as a contacts, windings etc.
To do so,
CPC 100
injects DC current into the test object, measures the current
that flows through the test object, and calculates the test object’s resistance.
The
Resistance
test card provides a total of three output ranges:
• 400A DC
• 6A DC
• V DC (2 wire)
The test setup depends on the selected range:
Figure 2:
Setup for a µ
Ω
measurement in the
400A DC range
Inject current from
400A DC
output to both sides of the test object. Input
VDC
measures the voltage drop at the test object, the current is measured internally
by
CPC 100
, and from these values the software calculates the test object’s
resistance.