CMC 356 Reference Manual
114
The
OSH256
connects to the EXIF socket of a
CMC 356
by means of the
adapter cable VEHK0010 (refer to section 9.6.5, "Adapter Cable for
Scanning Heads" on page 117) or a
CMLIB B
(refer to section 9.6.6,
For relay testing applications the
IFB256
serves to connect to a binary input
of a CMC test set (refer to section 9.6.2 below).
9.6.2 Interface
Box
IFB256
Typically, the combination of a scanning head
OSH256
with an interface box
IFB256
is used for relay testing when binary information (a trigger signal)
originates from a relay's LED. The
IFB256
is directly connected to the EXIF
socket of a
CMC 356
, which provides the auxiliary DC supply through the
IFB256
to the scanning head. The binary signal is connected to the inputs of
the CMC via banana plug leads.
Figure 9-12:
Interface box
IFB256
Order number: VEHZ1152
To OSH256
To binary input of the
CMC 356
To EXIF socket at the rear of the
CMC 356
Содержание CMC 356