Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 3 of 21
Smart Measurement Solutions
Smart Measurement Solutions
®
Figure 3: test signal shape - measurement settings
Figure 4: test signal shape - start continuous measurement
Expected Measurement result:
Figure 5: test signal shape - expected result
Waveform:
sinusoidal
V
peak-peak
:
~ 1.26 Volts (more accurate measurement follows)
Frequency: ~ 10 MHz (more accurate measurement follows)
If the waveform looks nice, the source amplifiers are working properly.