WMS-16 Weather Station Operator’s Manual
14
Sensor type Model Input Calibration
Wind Speed WS-01/02 P1
If you intend to add another sensor select A. If you do not want to add another sensor select Q.
You will be asked if you are sure that you want to erase memory. If you have added or deleted a sensor select Y
if not select N and press ENTER.
To initialize the Wind Direction sensor Press EXC to get the MAIN MENU.
From the MAIN MENU press number 1. STATION SET UP Then press ENTER
From the Station Set up menu press number 6. ADD OR REMOVE SENSORS. Then press Y, then ENTER.
From the Sensor Type Menu select number 3. Wind Direction. Then press ENTER
From the Wind Direction Sensor menu select number 1. (WS-02) then press ENTER.
From the Sensor Input Channel menu select number 11. ANALOG INPUT A7. Then Press ENTER.
The Sensor Configuration menu appears and it will show:
Sensor type Model Input Calibration
Wind Speed WS-01/02 P1
Wind Direction WS-02
A7
Temperature (thermistor) and Relative Humidity (capacitive sensor)
Installation
Connect the combination temperature and relative humidity sensor
=
s cable to the terminal strip on the back of
the control module. This sensor has a five conductor cable. The RED wire is connected to the +12 REF
terminal. The BLACK wire is connected to a GROUND terminal. The BROWN wire (humidity signal) is
connected to the A4 terminal. The WHITE (temperature signal) is connected to the A6 terminal and the GREEN
wire is connected to a GROUND terminal.
INITIALIZATION OF TEMPERATURE AND RELATIVE HUMIDITY
INITIALIZATION OF THE AUXILIARY TEMPERATURE SENSOR.
Press EXC to get the MAIN MENU.
From the MAIN MENU press number 1. STATION SET UP.