background image

Wiring 

The sensor is supplied with a 10 pin mating connector. When wired, it will contain power in, encoder 

(when optioned), and analog signal output connections.  Therefore, it is recommended that a cable that 
provides a shield around the analog signal wires be used.  For example, a cable that provides shielded pairs 
would suffice, so that the analog signal wires can be shielded from the power supply and encoder wires.  
Ground the shield at the DAQ system that is monitoring this sensor. 

The signal output is electrically isolated from the power supply.  This is done to prevent ground 

loops when the signal is fed into a single ended input.  This may cause trouble for a differential input due to 
common mode drift.  When using differential inputs, ground the negative input to an analog signal ground 
on the measuring DAQ system. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
Shunt Calibration 

When the shunt pins on the sensor connector are shorted, the sensor electronics will apply a shunt to the 
strain gage bridge, and shift the voltage output from it’s current position.  This is a real shunt of the bridge, 
not the star bridge shunt that so many rotary transformer systems use.  Reference the calibration sheet for the 
shunt calibration value.  Remove this short to return the sensor to normal operation.  

 
Cleaning: 

The TQ514 is a brushless sensor, and therefore, the only wearing parts are the bearings.  For the very low 
capacities, TQ514-003 and -006 units, these have oiled bearings and will periodically require a drop of high 
quality light machine oil on each.  Only a drop is required.  Clean environments will require a drop every 
360 straight hours of use.  Dirty or otherwise harsh environments may require more frequent oiling.  All 
other capacities have bearings that are permanently lubricated.  Do NOT immerse the assembly in the 
solvent.  This may damage the electronics inside the housing. 

 
Recalibration: 

To guarantee accuracy, all sensors should be recalibrated on an annual basis. 

Function 

Connector  Description 

+12-15Vdc 

Power Input 

+SIG 

Torque signal (nominally 

±

5V at fullscale)

 

-SIG (COM) 

Torque signal ground 

GND 

Power ground 

GND 

Encoder ground 

+5V 

Encoder power 

CW LEAD 

Encoder lead 

CW TRAIL 

Encoder trail 

GND 

Shunt calibration 

SHUNT 

Shunt calibration 

Содержание TQ514 Series

Страница 1: ...e mail info omega com For latest product manuals www omegamanual info TQ514 Series Digital Rotating Torque Sensors Shaft to Shaft Configuration TM Shop online at omega com User sGuide NORWALK CT...

Страница 2: ...pecifications without notice omega com info omega com Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Страница 3: ...meet the specifications and performance requirements described in the pamphlet Calibration was performed with a test system utilizing a reference load cell or deadweights and an electronic indicator...

Страница 4: ...200 in lb 363 0 31 6 259 0 216 0 17300 749 1875 TQ514 500 500 in lb 145 0 9 3 103 0 86 1 59000 TQ514 1K 1000 in lb 72 3 3 8 51 7 43 1 149000 TQ514 1 5K 1500 in lb 55 1 2 8 37 4 33 6 281000 TQ514 2K 20...

Страница 5: ...ent position This is a real shunt of the bridge not the star bridge shunt that so many rotary transformer systems use Reference the calibration sheet for the shunt calibration value Remove this short...

Страница 6: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Страница 7: ...s M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U...

Отзывы: