Omega TENS Скачать руководство пользователя страница 5

6

7

2.

Place the device 

over the desired treatment area.

Ensure that the electrodes are in contact with the skin.

3.

Press the power button on the Wireless 

host

Blue

lights will activate.

4.

Press the remote contol's power 

button

.

The light 

will be on and the device will be set to Mode 1

by default

. Th

is unit has 6 modes

Once 

the host 

 device

is 

successfully connected to the remote control

, the Wi-Fi

symbol will display on the remote control screen

.

 

If the

remote control 

is not operated or connected within two

minutes, it will automatically power down.

5.

Treatment will end on its own after 30 minutes

.

Treatment can be stopped anytime 

by pressing power

button on the 

host or remote control

DO NOT attempt to

remove the device from the body during treatment.

6.

Se

t the Mode

. Press the “M” button on remote

 control

 

to

set the mode

 or 

quickly tap the power button 

on

 

the host

device and then se

t the mode. For safety, whe

n the 

mode

is changed

, the intensity will automatically 

decrease to t

he

lowest setting.

 

To 

Change the

 

I

ntensity

:

 

On the host or

remote device, press t

he “+” 

to increase intensity or  the

”-”

button

 to decrease intensity.

7.

Time setting

Press remote's "T" button to set time.

Treatment can be set between 10-60mins in increments of 

10 mins

Pause Unit: 

Press 

the

       on remote control to 

pause 

and/or continue the treatment.

7.

To safely power off device during stimulation, you may

long press the       power button on the remote control or
press 

and hold the power button on the host device for 3

seconds

Replace protective film over electrodes to store

.

Note:

1. The

 

control 

unit

 is

 compatible with all Omega Wireless

Pads.

2.

When

 

host is used 

by itself

 the remote control 

can be

added 

mid-treatment

, and the strength will

 

reset to  the

default 

starting value after the remote is connected.

Unit Storage & Maintenance

Use a damp cloth with mild soap to gently wipe down

and clean the unit.

Note: 

This 

unit has

 water detection

 

technology. Do not

expose or

 submerge th

is unit

 in water or

 any

 other

 

liquids.

Water damage to the 

device or its components 

will void

the

 

warranty.

• Th

is unit 

should be operated, transported and stored

 

at

temperatures between 50° F and 104° F (10° C and 40°
C), with relative humidity between 30% - 85%. All values
have+/- 10% tolerance.
• Store the device, remote control

and 

electrodes

 in

original packaging 

between treatment sessions

.